|
|
Supplementary Comparison of Free-field Hydrophone Calibrations in the Frequency Range 250 Hz to 8 kHz |
CHEN Yi1,PING Zi-hong1,A.E.Isaev2,A.N.Matveev2 |
1.Hangzhou Applied Acoustics Research Institute, Fuyang, Zhejiang 311400, China
2. Russian National Research Institute for Physicotecnical and Radio Engineering Measurements,Mendeleeevo, Moscow region 141570, Russia |
|
|
Abstract A description is given of the results of COOMET.AUV.W-S1 supplementary comparison of free-field hydrophone calibration in the frequency range 250 Hz to 8 kHz between Hangzhou Applied Acoustics Research Institute – a pilot and Russian National Research Institute for Physicotecnical and Radio Engineering Measurements. Two standard hydrophones of TC 4033 and GI 55 were calibrated in this comparison. Reciprocity and comparison methods, their calibration facilities were used to assess the current state of free-field hydrophone calibration in low frequencies. The results indicate that the sensitivities of two hydrophones calibrated by China and Russia have good consistency.The deviations at most frequncies between the sensitivites and their reference values are less than 0.40 dB,the maximum deviation is 0.59 dB.The results prove that the lowest frequency of free-field hydrophone calibration of China and Russia can be extended to 250 Hz.
|
Received: 15 May 2014
Published: 10 December 2015
|
|
|
|
|
[1]BIPM. Mutual recognition of national measurement standards and of calibration and measurement certificates issued by national metrology institutes[S]. 1999.
[2]陈毅, Isaev A E, 王月兵, 等. 中俄在250Hz~200kHz频段的水听器比对校准[J].计量学报, 2012,33(1):47-51.
[3]Chen Yi, Isaev A E, Wang Yuebing, et al. COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 kHz[J]. Metrologia, 2011, 48, Tech Suppl, 09004.
[4]International Electrotechnical Commission. IEC 60565:2006. Underwater acoustics-Hydrophones-Calibration in the frequency range 0.01 Hz to 1 MHz[S]. 2006.
[5]Isaev A E, Matveev A N. Two approaches to hydrophone free-field calibration at continuous radiation in non-anechoic water tank[J]. Measurement Engineering, 2008, 12: 47-51.
[6]Isaev A E, Matveev A N. Use of a complex moving weighed averaging method for receiver non-uniform frequency response restoration [ J ]. Acoustical Physics, 2010, 56(5):693-696.
[7]Isaev A E, Matveev A N. Decreasing the free-field calibration error of hydrophones in a reverberating water tank [J].Measurement Techniques, 2013, 55(12):1451-1455.
[8]BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML. Guide to the expression of uncertainty in measurement[S].1995.
[9]Cox M G. The evaluation of key comparison data: determining the largest consistent subset[J]. Metrologia, 2005, 39(4): 589-595. |
|
|
|