Uncertainty Evaluation of Piezoelectric Constant Measurement Based on Dynamic Resonance Method |
ZHANG Rui-wen1,HE Long-biao2,ZHU Hai-jiang1,YANG Ping2 |
1. Beijing University of Chemical Technology, Beijing 100029, China;
2. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract A labview programming was written to achieve automatic measurements and data analysis. Based on the mathematical models established, the sweeping frequency step was optimized, and the characteristic parameters of piezoelectric ceramic NCE51 cylinders were tested and the uncertainty was evaluated. For 15mm thick sample, the relative expanded uncertainty of d33 measurement of piezoelectric ceramic is 0.96% (k=2), when the frequency step is selected as 50Hz. Thus the sample measured by dynamic resonance method can be used as reference material to calibrate the commercial d33 measurement apparatus, which uncertainty is about 2% to 5%.
|
|
|
|
|
[1]Fu J Y, Liu P Y, Cheng J P, et al. Optical measurement of the converse piezoelectric d33 coefficients of bulk and microtubular zinc oxide crystals[J]. Applied Physics Letters,2007,90 (21):1-3.
[2]Lamberti N, Iula A, Pappalardo M. The elecromechanical coupling factor in static and dynamic condition[J]. Acta Acustica united with Acustica,1999,85(1): 39-46.
[3]Hana P, Burianova L, Panos S. The dynamic method of determination of the piezoelectric hydrostatic coefficients[J]. Sensors and Actuators A: Physical,2004,110(1-3):318-322. [4]国家质量监督检疫检疫总局.GB/T 3389—2008 压电陶瓷材料性能测试方法-性能参数的测试[S].2008.
[5]Zhang Q, Qu M, Shen J G. Design and realization of piezoelectric crystal admittance circle measurement machine[J]. Piezoelectrics and Acoustooptics, 2005, 27(1):21-23,26.
[6]张沛霖, 张仲渊. 压电测量[M]. 北京:国防工业出版社, 1983, 112-430.
[7]He L B, Zhang R W, Zhu H J, et al. Measurement of piezoelectric strain constant based on virtual instrument technology by dynamic method[J]. Applied Mechanics and Materials, 2013, 333-335: 2422-2427. |
|
|
|