Determine the Content of Nucleic Bases Cytosine and Adenine by Electrode Potential Titration in Non-aqueous Solvent |
ZHANG Ling1, LIU Fang-fang2, CHEN Da-zhou2, WANG Jing1 |
1. National Institute of Metrology, Beijing 100013, China;
2. Beijing University of Chemical Technology, Beijing 100029, China |
|
|
Abstract A method to determine the content of nucleic bases Cytosine (Cyt) and Adenine (Ade) by electrode potential titration in non-aqueous soolvent is established.The factors affecting the experiments, such as non-aqueous solvent, titrant, the concentration of titrant, and end point of the titration are determined.The optimal experiment conditions determined are glacial acetic acid (HAc) as the non-aqueous solvent,perchloric acid as titrant (HClO4),titrant concentration 0.05 mol/L,and equivalence point of the electrode potential titration curve as endpoint.Through 15 times measurements and uncertainty evaluation, the purity of Ade is 99.40% ±0.38% (k=2), Cyt 99.58%±0.42% (k=2).
|
|
|
|
|
[1] Holden M J, Rabb S A, Tewari Y B, et al. Traceable phosphorus measurements by ICP-OES and HPLC for the quantitation of DNA[J]. Analytical Chemistry, 2007, 79(4):1536-1541.
[2] Yang I, Han M S, Yim Y H, et al. A strategy for establishing accurate quantitation standards of oligonucleotides: quantitation of phosphorus of DNA phosphodiester bonds using inductiVely coupled plasma-optical emission spectroscopy[J]. Anlytical Biochemistry, 2004,335(1):150-161.
[3] Kinumi T, Narukawa T, Fujii S, et al. Quantification of oligonucleotide containing a sequence failure product: comparison of isotope dilution mass spectrometry with other quantification methods[J]. European Journal of Mass Spectrometry,2009, 15(3): 399-407.
[4] O′Connor G, Dawson C, Woolford A, et al. Quantitation of oligonucleotides by phosphodiesterase digestion followed by isotope dilution mass spectrometry: Proof of concept[J].Analytical Chemistry,2002,74(15):3670-3676.
[5] Sigel H. Acid-base properties of purine residues and the effect of metal ions: Quantification of rare nucleobase tautomers[J].Pure Appl Chem, 2004, 76(10):1869–1886.
[6] 梁忆,张皋,陈智群,等.非水溶剂电位滴定法测定标准物质RDX和HMX的含量[J].火炸药学报,2008, 31(2):105-108.
[7] 梁忆,胡玲,张皋,等.标准物质RDX、HMX定值的不确定度评估[J].化学分析计量,2007,16(5): 4-6.
[8] JJG1006—1994,一级标准物质[s].
[9] 全浩,韩永志.标准物质及其应用[M].(第2版).北京:中国标准出版社,2003. |
|
|
|