Study on Peak Power Meter Calibration |
ZHAO Hai-ning,HOU Li-xin,ZHOU Xin |
National Institute of Metrology, Beijing 100013, China |
|
|
Abstract In modern digital mobile communications system, the signals are transmitted in burst mode that requires power meter be able to measure the power of discontinuous signal, so frist,the peak power and burst power meters are introduced. Those special parameters about the power of burst signal and the measurement theory of peak power meter are also introduced. The theory and method of calibrating peak power meter are described combined with the characters of those parameters. Finally, calibration results and measurement uncertainty analysis are reported,the expanded uncertainty is 0.17 dB, the system can meet the calibration requirements of the peak power meter.
|
|
|
|
|
|
[1]Anderson A B. Measuring Power Levels In Modern Communications Systems [J]. MICROWAVES & RF,2000,(10):89-98.
[2]Agilent Technologies.Fundamentals of RF and Microwave Power Measurements[Z].Application Note 64-1C,2001.
[3]Agilent Technologies.4 Steps for Better Power Measurements[Z]. Application Note 64-4C,2000.
[4]JJG 2009—1987,射频与微波功率计量器具检定系统[S].
[5]吕品,高飞,崔孝海.基于“数据驱动计量”的通用微波小功率自动计量系统[J].计量学报,2007,28(1):77~80.
[6]JJF 1059—1999,测量不确定度评定与表示[S]. |
|
|
|