Study on Static Balance Measurement of Gyroscope |
LIU Guo-dong1,LIU Bing-guo1,LI Jia2,ZHU Xiao-fang1,ZHANG Zhuo1 |
1. Institute of Optical Instrument & Automation, Harbin Institute of Technology, Harbin,Heilongjiang 150001, China
2. Beijing Xinghua Machinery Factory, Beijing 100854, China |
|
|
Abstract A novel design of the rotor static balance measuring instrument structure is introduced, in which an agate blade bearing and autocollimator are adopted, and high accuracy measurement for rotor static balance is realized. A novel mechanism combining coarse fine double blades bearing ensures high precision and wide range, reduces the wear and tear of blades, and improves the efficiency of measurement. The sensitivity coefficient of the instrument is adjusted by measuring the swing period of rotor bearer. The instrument is calibrated by standard weights, experiment results show that the precision of the instrument is about 4mg·mm and the measurement range is about 2000 mg·mm.
|
|
|
|
|
|
[1]李丙乐,王厚生.陀螺转子的称重静平衡实验研究[J].工程设计学报,2006,13(5):332-335.
[2]卢志辉,薄悦,张磊乐, 等. 一种新型的质心测量机构与测量方法[J]. 计量学报,2010, 31 (2):119-122.
[3]姚国年, 卫军胡, 赵希林, 孙国基. 高精度弹丸质偏心测试系统及其误差分析[J]. 仪器仪表学报, 2009, 30(14): 862-866.
[4]郑勇, 常晓东, 崔瑞禧. 战斗部径向质心测量技术研究[J]. 航天制造技术, 2009,12(6):49-51.
[5]张琳,杨涛,王佳民.半液浮陀螺浮子的静平衡方法研究[J].弹箭与制导学报,2006,27(3):74-76.
[6]王民学, 洪沅芷, 施乐平, 等. 高精度垂直轴微力矩标定仪的研制[J]. 计量学报,2008,29(3): 231-234.
[7]郑红梅, 刘正士, 郑传荣, 等. 机器人六维腕力传感器标定试验台误差分析与研究[J]. 计量学报,2005,26(4): 333-336. |
|
|
|