A Coaxial-to-waveguide Adaptor Calibration Method using an Adjustable Shorter with Unspecified Phases
WU Chang-ying,XIE Chun-jian,WEI Gao,XU Jia-dong
School of Electronic Information, Northwestern Polytechnical University Chang'an Campus, Xian, Shaanxi 710129, China
Download: PDF (863 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  A technique is described to remove the coaxial-to-waveguide adaptor errors when the reflection coefficient of waveguide device is measured by vector network analyzer.This technique utilizes a matched load and an adjustable shorter with several unspecified phases.And one of the shorted planes is used as reference plane.It is shown that the technique is resistant to the error of the adjustable shorter position and is easy to implement.The efficiency of this technique is verified by measurement data in X band.
Key wordsMetrolog      Coaxial-to-waveguide adaptor      Adjustable shorter      Matched load      Calibration     
PACS:  TB973  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
WU Chang-ying
XIE Chun-jian
WEI Gao
XU Jia-dong
Cite this article:   
WU Chang-ying,XIE Chun-jian,WEI Gao, et al. A Coaxial-to-waveguide Adaptor Calibration Method using an Adjustable Shorter with Unspecified Phases[J]. Acta Metrologica Sinica, 2011, 32(2): 168-171.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2011.02.16     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2011/V32/I2/168
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech