According to the needs of calibrating the probing size error, a method based on the standard balls for the calibration of industrial CT probing size error is proposed and factors which cause the probing size error are detailly analyzed. Firstly, it calibrates the diameter by the length-measuring machine of National Institute of Metrology, then calibrates the industrial CT probing size error by these balls. The experimental results show that these standard balls can be calibrated accurately, and high quality image with lower artifacts can be acquired. And this kind of standards can meet the calibration needs of probing size error of industrial CT system. Error sources in imaging process, which influence the probing size error are discussed and analyzed. It is of great significance to establish calibration and traceability method for industrial CT system.
王义旭, 施玉书, 高思田, 李东升, 李伟, 李琪, 宋旭. 工业CT探测尺寸误差的校准及误差分析[J]. 计量学报, 2014, 35(3): 216-220.
WANG Yi-xu, SHI Yu-shu, GAO Si-tian, LI Dong-sheng, LI Wei, LI Qi, SONG Xu. Calibration and Analysis for Probing Size Error of Industrial CT. Acta Metrologica Sinica, 2014, 35(3): 216-220.
[1] Bartscher M, Neuschaefer-Rube U, Wldele F. Computed tomography—a highly potential tool for industrial quality control and production near measurements [J]. CIRP Annals-Manufacturing Technology, 2004,(1860):477-796.
[2] Bartscher M, Neuschaefer-Rube U, Staude A, et al. Application of an industrial CT reference standard for cast free-form shaped work pieces[C]// International Symposium on Digital Industrial Radiology and Computed Tomography. Berlin, Germany, 2011, 20-22.
[3] Voet A, Kiekens K, Welkenhuyzen F, et al. A test object for calibration and accuracy assessment in X-ray CT metrology[C] // 10th International Symposium on Measurement and Quality Control 2010. Osaka, Japan, 2010.
[4] 陈思,陈浩,李敬.一种工业CT测量精度评估方法[C]//全国射线数字成像与CT新技术.四川绵阳,2012,189-197.
[5] Brunke O, Suppes A. Precise 3D dimensional metrology using high resolution X-ray computed tomography (μCT)[C]// SPIE. San Diego, California, 2010, (7804): 780400-1.
[6] Prufstelle Testing and Certification Institute.VDI/VDE 2630-1.3[S]. Computed Tomography in dimensional measurement.2011.
[7] 倪育才.实用测量不确定度评定[M].北京:中国计量出版社, 2009, 125-126.
[8] 朱训生,任新生,杨建国.球度最小二乘评定的一般公式[J].上海交通大学学报,1998,32(5):43-45.
[9] 张朝宗, 郭志平.工业CT技术和原理[M].北京:科学出版社, 2009, 35-41.
[10] 杨倩,潘晋孝.工业CT中机械定位误差对投影数据的影响[J].测试技术学报, 2009, 23(增刊):138-143.
[11] 王钰,陈教泽.工业CT探测系统噪声特性研究[J]. 核电子学与探测技术, 2010, 30(7):929-934.
[12] 张峰,闫镔.工业X-CT散射校正技术综述[J].CT理论与应用研究, 2009, 18(4):34-43.
[13] 刘晓鹏.锥束工业CT图像校正技术研究[D].西安:西北工业大学, 2006.
[14] 王京东, 夏顺仁. 图像重建算法评价和误差分析的研究进展[J].国外医学生物医学工程分册, 1998, 21(6):332-337.