1. National Institute of Metrology, Beijing 100029, China
2. Harbin Institute of Technology, Harbin, Heilongjiang 150001, China
3. Beijing Institute of Technology, Beijing 100081, China
Abstract:According to the definition of contrast and contrast sensitivity function, a method of producing test targets for human eye contrast sensitivity measurement is proposed with integrating spheres and special light modulators. Then an automated measurement system is built and human eye contrast sensitivity function can be measured by psychophysical experiment with cross ladder method. Its spatial frequency covers 0.8, 1.5, 2.0, 3.0, 4.5, 6.0, 12.0 and 18.0 cycles per degree of visual angle. The indication error of background luminance of test targets satisfies ±5%, and contrast repeatability is 0.001 and reproducibility is 0.004, which meet requirements of human eye contrast sensitivity measurement.
[1]Schad O H. Optical and photoelectric analog of the eye [J]. Journal of the Optical Society of America, 1956, 46(9):721-739.
[2]谢培英,迟蕙. 眼视光医学检查和验配程序[M]. 北京:北京大学医学出版社,2006.
[3]Wooten B R, Renzi L M, Moore R, et al. A practical method of measuring the human temporal contrast sensitivity function [J]. Biomedical Optics Express, 2010, 1(1):47-58.
[4]邓述移, 徐智敏, 杜连芳, 等. 人眼调制传递函数[J]. 兵工学报, 1982, 3(2):24-33.
[5]姚军财, 石俊生, 杨卫平, 等. 人眼对比度敏感视觉特性及模型研究[J]. 光学技术, 2009, 35(3):334-337.
[6]石俊生, 姚军财, 余鸿飞, 等. 用显示器测量人眼衬比度敏感函数[J]. 光学学报, 2007, 27(4):744-748.
[7]张吉焱, 定翔, 钟波. 视听诊断设备计量标准研究[J]. 中国计量, 2011, (12):23-24.
[8]张吉焱,李姣,高明亮, 等. 角膜曲率计轴位标准器的研制[J]. 计量学报, 2017, 38(3): 276-279.