Abstract:The radiation equipment for air-kerma primary standard in the range of low-energy X-rays is established.The results of experiment show that the diameter of the radiation field is more than 40 mm within 1% non-uniformity.The distance from the source to the reference plane is 1 000 mm.Radiation qualities recommend by CCRI are established.The half value layer(HVL)and the homogeneity coefficient are obtained by measuring the curve of x-rays attenuation.This low-energy X-rays radiation equipment and research results are the important base of establishing primary standard,realization of air-kerma and the APMP key comparison in the range of low-energy X-rays(APMP.RI(I)-K2).
吴金杰,王培玮,段小娟, 李兵,杨元第. 低能X射线基准辐射装置的建立[J]. 计量学报, 2011, 32(z1): 9-13.
WU Jin-jie,WANG Pei-wei,DUAN Xiao-juan,LI Bing,YANG Yuan-di. The Establishment of Radiation Primary Standard in the Range of Low Energy X-rays. Acta Metrologica Sinica, 2011, 32(z1): 9-13.
[1]Burns D T,Degree of equivalence for the key comparison BIPM.RI(I)-K2 between national primary standards for low-energy x-rays[J].Metrologia,2003,44(Tech.Suppl.)06006:1-15.
[2]NIMJ,Technical protocol for APMP key comparison of air kerma standards for the CCRI reference qualities and the ISO 4037 narrow spectrum series in the low-energy x-ray region[EB],http://kcdb.bipm.org/,2010.
[3]ISO 4037-1, X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy-Part1: Radiation characteristics and production methods[S].1996.
[4]BIPM,Qualities de rayonnement [R],CCEMRI(I),1972.