2025年05月13日 星期二 首页   |    期刊介绍   |    编 委 会   |    投稿指南   |    期刊订阅   |    统合信息   |    联系我们
计量学报  2011, Vol. 32 Issue (2): 123-125    DOI: 10.3969/j.issn.1000-1158.2011.02.07
  本期目录 | 过刊浏览 | 高级检索 |
CCD自准直仪示值漂移的自校准系统
张俊杰1,2,王震2,李政阳2,张俊秀2
1. 天津大学 精密仪器与光电子工程学院国家重点实验室, 天津 300072
2. 北京航天计量测试技术研究所, 北京 100076
Self-calibration System for the Value Drifts of CCD Autocollimator
ZHANG Jun-jie1,2,WANG Zhen2,LI Zheng-yang2,ZHANG Jun-xiu2
1. State Key Lab of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China
2. Beijing Aerospace Institute for Metrology and Measurement Technology, Beijing 100076, China
全文: PDF (558 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 数显式自准直仪使用时,气流等环境条件的影响以及仪器本身发热等原因都会引起示值的漂移和跳动,使测量误差增大、测量结果不能回零。采用增加一块固定反光镜作为参考镜,它与工作反光镜以自准直仪进行差动,组成自校准系统,可以有效地进行消减。仅用工作反光镜时2h内的漂移量为0.03″,经自校准系统后为0.01″;样本标准偏差为0.091 1″,经自校准后为0.057 4″。差动式反光镜自校准系统可以显著减小CCD自准直仪的示值漂移,对减小示值跳动也具有一定作用。且无论是对环境条件还是仪器本身原因引起的漂移和跳动都是有效的。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
张俊杰
王震
李政阳
张俊秀
关键词 计量学差动法自准直仪示值漂移自校准    
Abstract:When the digital autocollimator is put into use directly, its indications will drift and beat under the influence of environmental conditions such as air streams and the increasing heat of equipment itself, which leads to the measuring errors greatly increased and the measurement results could not return to zero. A fixed mirror which is added as a reference mirror makes the differential measurement with a working mirror and forms a self-calibration system, which could effectively reduce errors. The value of drift is 0.03″ in two hours when working mirror is only used and it reduces to 0.01″ after the adoption of a self-calibration system; the standard deviation of samples is 0.0911″ and it reduces to 0.0574″ after the self-calibration. The self-calibration system of differential mirror could effectively reduce the indication errors of CCD autocollimator and plays an efficient role in reducing the indication beat. It is an effective way for the system to reduce the drift and beat of the equipment caused by both the environmental conditions and the equipment themselves.
Key wordsMetrology    Differential method    Autocollimator    Indication drifts    Self-calibration
    
PACS:  TB92  
作者简介: 张俊杰(1982-),男,山西原平人,天津大学博士研究生,研究方向为几何量精密仪器、图像处理等。joojay@gmail.com
引用本文:   
张俊杰,王震,李政阳,张俊秀. CCD自准直仪示值漂移的自校准系统[J]. 计量学报, 2011, 32(2): 123-125.
ZHANG Jun-jie,WANG Zhen,LI Zheng-yang,ZHANG Jun-xiu. Self-calibration System for the Value Drifts of CCD Autocollimator. Acta Metrologica Sinica, 2011, 32(2): 123-125.
链接本文:  
http://jlxb.china-csm.org:81/Jwk_jlxb/CN/10.3969/j.issn.1000-1158.2011.02.07     或     http://jlxb.china-csm.org:81/Jwk_jlxb/CN/Y2011/V32/I2/123
京ICP备:14006989号-1
版权所有 © 《计量学报》编辑部
地址:北三环东路18号(北京1413信箱)  邮编:100029 电话:(010)64271480
本系统由北京玛格泰克科技发展有限公司设计开发  技术支持:support@magtech.com.cn