1.中国计量科学研究院, 北京 100029
2. All Russian Research Institute for Optical and Physical Measurements, Ozernaya 46, Moscow 119361, Russia
Bilateral Comparison of Spectral Irradiance between NIM and VNIIOFI
DAI Cai-hong1,WU Zhi-feng1,WANG Yan-fei1,LIN Yan-dong1,Boris Khlevnoy2
1. National Institute of Metrology, Beijing 100029, China
2. All Russian Research Institute for Optical and Physical Measurements (VNIIOFI), Ozernaya 46, Moscow 119361, Russia
Abstract:With the new spectral irradiance measurement facility based on blackbody BB3500M of NIM (National Institute of Metrology), a bilateral spectral irradiance comparison was carried out between NIM and VNIIOFI (All Russian Research Institute for Optical and Physical Measurements) in the spectral wavelength from 250 nm to 2500 nm in the period from December 2014 to January 2015. The temperature measurement of the high temperature blackbodies were traced to the Pt-C and Re-C fixed point blackbodies and checked against the WC-C fixed-point blackbody for the two institutes respectively. The consistency of the temperature at 3021K is better than 70mK. The comparison result shows that the average relative deviation of spectral irradiance at the 44 wavelengths is 0.45%. The consistency is better than 0.9% except the maximum deviation 1.1% at the wavelength of 2000nm. The spectral irradiance units measured by NIM and VNIIOFI in this comparison are in agreement within the combined standard uncertainties of the laboratories over the wavelength range compared.
[1]代彩红,吴志峰,欧阳慧泉,等. 第四代光谱辐射度和色温度国家基准装置的研制[J].计量学报,2013,34(3):201-206.
[2]DAI C H, WU Z F, YU J L. Realization of the new national primary scale of spectral radiance and spectral irradiance[C]//Proc SPIE 8910, International Symposium on Photoelectronic Detection and Imaging 2013: 89100D, 2013.
[3]DAI C H, WU Z F, YU J L, et al. Investigation of Blackbody 3500M[C]//Proc SPIE 8192, International Symposium on Photoelectronic Detection and Imaging: 819250, 2011.
[4]WU Z F, DAI C H, YU J L, et al. Realization of the 250nm-400nm spectral irradiance scale [C] //Proc SPIE 8201, 2011 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems, 82012B, 2011.
[5]OUYNG H Q, DAI C H, HANG B, et al. Bandwidth determination of a Czerny-Turner double monochromator with varied slit widths through experiment and data computation[C]//Proc SPIE, 81971B, 2011.
[6]Woolliams E R, Harrison N J, Fox N P. Preliminary results of the investigation of a 3500K black body[J]. Metrologia, 2000, 37: 501-504.
[7]Sperfeld P, Metzdorf J, Harrison N J, et al. Investigation of high-temperature black body BB3200[J]. Metrologia, 1998, 35: 419-422.
[8]Woolliams E R, Fox N P, Maurice G Cox, et al. The CCPR K1-a key comparison of spectral irradiance from 250 nm to 2500 nm: measurements, analysis and results[J]. Metrologia, 2006,43(2): S98-S104.