Abstract:When using photometer with nonuniform responsivity distribution to measure the averaged LED intensity of LED with nonuniform illuminance distribution, the measurement results will have significant errors. A special experiment system is designed to measure the responsivity distribution of three types of photodetectors, and the measurement results indicated that the responsivity uniformity of the photodetector with diffuser is worse and the corresponding measurement error can be 2.0% when it is used to measure a white LED's averaged intensity. The nonuniformity evaluation method and the expression of evaluation factor c are proposed. It is suggested that the photodetectors with diffuser should be avoid when calibrating the averaged LED intensity.
刘建,刘慧,赵伟强,杨臣铸. 探测器响应度不均匀性对测量LED平均发光强度的影响[J]. 计量学报, 2012, 33(4): 317-320.
LIU Jian,LIU Hui,ZHAO Wei-qiang,YANG Chen-zhu. The Impact of Uniformity of Detector Responsivity on Measurement of Averaged LED Intensity. Acta Metrologica Sinica, 2012, 33(4): 317-320.
[1]赵伟强,刘慧.脉冲与直流电源供电的LED光电特性的比较分析[J].计量学报,2009,30(6A):163-166.
[2]刘慧,杨臣铸.光度测量技术[M].中国计量出版社,2011.
[3]代彩红,于家琳,于靖,等.球冠形入射光学系统中等效辐射平面位置的计算[J].计量学报,2006,27(3):236-240.
[4]CIE.MEASUREMENT OF LED[R].CIE127-2007.
[5]林延东,姚和军.陷阱探测器面响应均匀性的测量[J].现代计量测试,2000,(3):17-21.
[6]代彩红,吕正,于家琳,等.基于物理模型的光谱辐射照度标准灯数据处理方法[J].计量学报,2009,30(1):6-10.
[7]熊利民,李平,王煜,等.用数值模拟方法配置光度探测器[J].计量学报,2006,27(2):121-123.
[8]Gong H, Hanssen L M, Eppeldauer G P. Spatial and angular responsivity measurements of photoconductive HgCdTe LWIR radiometers[J]. Metrologia, 2004, 41(3): 161-166.
[9]李小亭,胡金敏.计量光学[M].北京:中国计量出版社,2003.
[10]Yoshihiro O. Photometric Calibrations[R]. NIST Special Publication 250-37, 1997.