Abstract:Feature length of acoustic resonator cavity is one of the important parameters when measuring Boltzmann constant with acoustic method. Its measurement uncertainty directly affects the uncertainty of Boltzmann constant k. The development of high-precision measurement system for feature length of acoustic resonance cavity is introduced, its key technology described, and the measurement result with a real time correcting about the environmental parameters is achieved. According to the test data, this measurement system can get the uncertainty of measurement of 0.60μm, standard deviation of 0.20μm in 3 times measurement.
[1]Fellmuth B, Gaiser Ch, Fischer J. Determination of the Boltzmann constant:status and prospects[J]. Measurement Science and Technology,2006,17(10):R145-R159.
[2]段宇宁,林鸿,张金涛,等.基准声学温度计的研究进展[J].计量学报, 2007, 28(3A): 1-6.
[3]何学明,陈学东,曾理湛, 等.超精密气浮平台的定位精度分析[J].华中科技大学学报(自然科学版),2008,36(3):8-11.