Abstract:A digital compensator is designed to output regulated signal in four-quadrant and its core component is two 16-bit serial input multiplying digital-to-analog converters. With sinusoidal reference signal, the in-phase and the quadrature component of the output can be adjusted in accordance with the given code. A computer controlled current-comparator bridge is developed and magnetic potential self-balance can be got by applying the digital compensator and Newton iteration algorithm. The AC/DC relative difference of bifilar helix resistors are measured below 1592 Hz through the proposed bridge using substitution method, and the standard measurement uncertainty is 2×10-7 that can meet the requirement for quantity value dissemination of AC resistance.
王云静,王玉田,曲正伟. 数字微差补偿器及其在电流比较仪电桥中的应用[J]. 计量学报, 2015, 36(5): 530-534.
WANG Yun-jing,WANG Yu-tian,QU Zheng-wei. A Digital Compensator and Its Application in Current-comparator Bridge. Acta Metrologica Sinica, 2015, 36(5): 530-534.
[1]黄璐, 张钟华, 赵伟. 交流电阻标准研究进展[J]. 电测与仪表, 2004, 41(6):1-4.
[2]Li Z K, He Q, Huang L, et al. A new type AC resistor with calculable frequency dependency[J]. 计量学报, 2008, 29(1): 241-247.
[3]曲正伟, 张钟华, 赵伟. 双螺线电阻静电屏蔽方法研究[J]. 仪器仪表学报,2010, 31(2): 476-480.
[4]张钟华, 贺青, 李正坤, 等. 量子化霍尔电阻国家标准的研究[J]. 计量学报, 2005, 26(2):97-101.
[5]张钟华. 量子计量基准现状[J]. 仪器仪表学报,2011, 32(1):1-5.
[6]李正坤, 张钟华, 贺青, 等. 低温电流比较仪动态特性的研究[J]. 仪器仪表学报,2006, 27(8):825-829.
[7]黄璐, 张钟华, 赵伟, 等. 精密宽频开尔文电桥的研制[J]. 计量学报,2008, 29(2):97-101.
[8]Lu W J, Yang Y, Wang W, et al. Improvements of AC/DC calculable standard resistor with double helix arrangement and Kelvin resistance bridge at NIM[C]//CPEM2010: Conference on Precision Electromagnetic Measurements, Korea, Daejeon, 2010.
[9]Jeffery A, Shields J Q, Shields S H. A multi-frequency 4-terminal-pair AC bridge[C]//CPEM2000 Conference Digest, Australia, Sydney, 2000.
[10]曲正伟, 赵伟, 李正坤. 交流电流比较仪在四端钮阻抗测量中的应用[J]. 电测与仪表, 2009, 46(12): 1-5.
[11]曲正伟, 王云静. 基于电流比较仪的四端钮阻抗电桥[J]. 仪器仪表学报,2011, 32(9): 1987-1992.
[12]Callegaro L. On strategies for automatic bridge balancing[J].Instrumentation and Measurement,2005, 54(2): 529-532.