Abstract:The principle and process of the measurement are presented. The experimental results is analysed,the influence of vacuum on the transmittance of Brewster window is negligible compared with the uncertainty of optical power measurement and reproducibility of window transmittance measurement. The window transmittance measured in the air meet the requirement of cryogenic radiometer.
白山,林延东,吕亮. 真空状态对布儒斯特窗透射比影响的研究[J]. 计量学报, 2013, 34(6): 529-531.
BAI Shan,LIN Yan-dong,LV Liang. Investigation of the Influence of Vacuum on the Transmittance of Brewster Window. Acta Metrologica Sinica, 2013, 34(6): 529-531.
[1]Quinn T J. Primary methods of measurement and primary standards [J]. Metrologia, 1997,34:61-65.
[2]林延东.计量科学研究50年[M].北京:中国计量出版社,2004,839-845.
[3]Krh P. Toivanen P, Manoochehri F, et al. Development of a detector-based absolute spectral irradiance scale in the 380—900 nm spectral range[J]. Applied Optics, 1997, 36: 8909-8918.
[4]Goebel R, Yilmaz S. Stability under vacuum of silicon trap detectors and their use as transfer instruments in cryogenic radiometry[J]. Applied Optics, 1996,35(22):4404-4407.
[5]Fox N P. Trap detectors and their properties[J]. Metrologia,1991, 28: 197-202.
[6]林延东,吕亮,李奇.低温辐射计光辐射有效加热功率测量的研究[J].计量学报,2009,30(6A):26-30.