2025年04月06日 星期日 首页   |    期刊介绍   |    编 委 会   |    投稿指南   |    期刊订阅   |    统合信息   |    联系我们
计量学报  2024, Vol. 45 Issue (10): 1570-1575    DOI: 10.3969/j.issn.1000-1158.2024.10.19
  电离辐射、化学计量与生物计量 本期目录 | 过刊浏览 | 高级检索 |
利用数字化仪实现放射性气体85Kr核素活度浓度的绝对测量
拓飞1,路天翔2,张明3,周倩倩3,郭贵银4,王硕1
1.中国疾病预防控制中心辐射防护与核安全医学所,北京100088
2.国家卫生健康委职业安全卫生研究中心,北京102308
3.中国计量科学研究院,北京100029
4.苏州热工研究院,江苏 苏州215004
Absolute Measurement of Radioactive Gas 85Kr Activity Concentration Using a Digitizer
TUO Fei1,LU Tianxiang2,ZHANG Ming3,ZHOU Qianqian3,GUO Guiyin4,WANG Shuo1
1.National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088,China
2.National Center for Occupational Safety and Health, NHC, Beijing 102308,China
3.National Institute of Metrology, Beijing 100029, China
4.Suzhou Nuclear Power Research Institute Co., Ltd, Suzhou, Jiangsu 215004, China
全文: PDF (499 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 基于中国计量科学研究院建立的长度补偿法充气正比计数器绝对测量系统,利用CAEN公司开发的DT5730型数字化仪替代多套传统模拟电子学插件, 并通过数字化仪对β信号进行采集、重建和数据离线分析,开展85Kr放射性惰性气体活度浓度的绝对测量。实验研究了各种条件的影响,通过对正比计数器端效应、正比计数器壁效应、本底、噪声、死时间漏计数和阈值以下计数进行修正,最终得到85Kr活度浓度绝对测量结果。结果表示:在相同阈值、主放大倍数100倍、气压50.5kPa,工作电压1675V和1700V的条件下,数字化仪测量和电子学插件测量85Kr放射性活度浓度的相对偏差分别为0.83%和0.74%。测量结果在不确定度范围内一致,证明数字化仪可有效替代传统电子学插件实现放射性惰性气体85Kr活度浓度的绝对测量。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
拓飞
路天翔
张明
周倩倩
郭贵银
王硕
关键词 电离辐射计量数字化仪85Kr正比计数器放射性活度浓度绝对测量    
Abstract:Based on the length compensation method absolute measurement system for gas-filled proportional counters established by the National Institute of Metrology, this study utilizes the DT5730 digital analyzer developed by CAEN to replace multiple traditional analog electronic modules. The digital analyzer is employed to collect, reconstruct, and analyze β signals offline to perform the absolute measurement of the activity concentration of 85Kr radioactive inert gas.The influence of various conditions is experimentally studied, and the absolute measurement results of 85Kr activity concentration are obtained by correcting the proportional counter end effect, the proportional counter wall effect, background, noise, dead time leakage counts, and sub-threshold counts.Under the same threshold, with a main magnification factor of 100 times, gas pressure of 50.5kPa, and operating voltages of 1675V and 1700V, the relative deviations of 85Kr radioactive activity concentration measurements between the digital analyzer and the electronic modules are 0.83% and 0.74%, respectively. The measurement results are consistent within the uncertainty range, proving that the digital analyzer can effectively replace traditional electronic modules for the absolute measurement of the 85Kr activity concentration.
Key wordsionizing radiation measurement    digitizer    85Kr;proportional counter;radioactive activity concentration;absolute measurement
收稿日期: 2023-09-26      发布日期: 2024-09-30
PACS:  TB98  
通讯作者: 张明(1984-),男,吉林辽源人,中国计量科学研究院副研究员,从事电离辐射计量方面的研究。Email:zhming@nim.ac.cn     E-mail: zhming@nim.ac.cn
作者简介: 拓飞(1980-),男,甘肃庆阳人,中国疾病预防控制中心辐射防护与核安全医学所研究员,从事辐射检测与评价方面的工作。Email:tuofei@nirp.chinacdc.cn
引用本文:   
拓飞,路天翔,张明,周倩倩,郭贵银,王硕. 利用数字化仪实现放射性气体85Kr核素活度浓度的绝对测量[J]. 计量学报, 2024, 45(10): 1570-1575.
TUO Fei,LU Tianxiang,ZHANG Ming,ZHOU Qianqian,GUO Guiyin,WANG Shuo. Absolute Measurement of Radioactive Gas 85Kr Activity Concentration Using a Digitizer. Acta Metrologica Sinica, 2024, 45(10): 1570-1575.
链接本文:  
http://jlxb.china-csm.org:81/Jwk_jlxb/CN/10.3969/j.issn.1000-1158.2024.10.19     或     http://jlxb.china-csm.org:81/Jwk_jlxb/CN/Y2024/V45/I10/1570
京ICP备:14006989号-1
版权所有 © 《计量学报》编辑部
地址:北三环东路18号(北京1413信箱)  邮编:100029 电话:(010)64271480
本系统由北京玛格泰克科技发展有限公司设计开发  技术支持:support@magtech.com.cn