Abstract:The distortion of engine intake temperature affects the stability of the engine, and small inertia thermocouples are usually used to measure its rapidly changing airflow temperature. In order to accurately and comprehensively obtain the dynamic characteristic of small inertia thermocouples, a dynamic characteristic calibration and evaluation method of small inertia thermocouples is porposed, and the construction of dynamic calibration device, calibration test and frequency domain analysis are introduced. The dynamic calibration method is combined with a pulse laser and a calibration wind tunnel, through the analysis of amplitude frequency characteristics and phase frequency characteristics, the evaluation method for the dynamic characteristics of small inertia thermocouples is expanded. Using this calibration method for dynamic calibration of small inertia thermocouples, the maximum relative uncertainty of type A reduces from 7.2% to 2.4%, which can be used for dynamic calibration of fast response temperature sensors below 30ms.It provides technical reference for high-frequency dynamic temperature measurement such as engine intake temperature distortion.
GUMEN V, ILLYAS B, MAQSOOD A, et al. High-Temperature Thermal Conductivity of Ceramic Fibers[J]. Journal of Materials Engineering and Performance, 2001, 10: 475-478.
ERELL E, LEAL V, MALDONADO E. Measurement of air temperature in the presence of a large radiant flux: an assessment of passively ventilated thermometer screens[J]. Boundery-Layer Meteorology, 2005, 114: 205-231.
[4]
STRUK P, DIETRICH D, VALENTINE R, et al. Comparisons of gas-phase temperature measurements in a flame using thin-filament pyrometry and thermocouples[C]//41st Aerospace Sciences Meeting and Exhibit. Reno, US, 2003.
LI C Y. Directional Sensitivity of Dynamic Characteristic of Total Temperature Sensor [J]. Aviation Metrology & Measurement Technology, 2008, 28(1): 2-3.
ZENG X H, QI L H, XIAO Y, et al. Droplet temperature measurement and calculation during spray process[J]. Chinese Journal of Scientific Instrument, 2011, 32(1): 93-98.
CUI Y X, XUE S Y, ZHOU T, et al. Fabrication and performance analysis of thin film transient temperature sensor[J]. Chinese Journal of Scientific Instrument, 2017, 38(12): 3028-3035.
ZHAO J. Design Method of high gas temperature sensorbased on double heat conduction equation[J]. ACTA Metrologica Sinica, 2022, 43(2): 215-221.
[16]
ITO Y, INOKURA N, NAGASAKI T. Conjugate heat transfer in air-to-refrigerant airfoil heat exchangers[J]. ASME Journal of Transfer, 2014, 136: 1-12.
[18]
ZHENG C H, CHENG L M, SAVELIEV A, et al. Gas and solid phase temperature measurements of orous media combustion[J]. Proceedings of the Combustion Institute, 2011, 33(2): 3301-3308.
[2]
KNAUSS H, GAISBAUER U, WAGNER S, et al. Calibration experiments of a new active fast response heat flux sensor to measure total temperature fluctuations[C]//Intern Conf on the Methods of Aerophys Research. 2002.
[5]
CHERNOVSKY M K, ATREYA A. Transient Measurements of Temperature and Radiation Intensity in Spherical Microgravity Diffusion Flames[C]//44st AIAA Aerospace Sciences Meeting and Exhibit. Reno, US, 2006.
YANG Z X, GU Z H, ZHANG W Q, et al. The error compensation method of the low-speed wind tunnel flow temperature based on the thermocouple[J]. Chinese Journal of Scientific Instrument, 2022, 43(5): 68-76.
DUKHAN N, AL-RAMMAHI M A, SULEIMAN A S. Fluid temperature measurements inside metal foam and comparison to Brinkman-Darcy flow convection analysis[J]. International Journal of Heat and Mass Transfer, 2013, 67: 877-884.
ZHAN Z H, NING F, NING J X. Study on some measuring technologies of temperature distortion on inlet flow of engine[J]. Journal of Data Acquisition & Processing, 2000, 15(2): 226-229.
LI H Y, ZHAO J. The research of dynamic calibration for temperature sensors[J]. Metrology& Measurement Technology, 2008, 28(1): 2-3.
YANG Z X, GU Z H, ZENG X, et al. Study on dynamic characteristics evaluation method of the cascade thermocouple system[J]. Chinese Journal of Scientific Instrument, 2020, 41(11): 74-81.
WU P, LIN T. Research on identification modeling of sheathed thermocouple sensor based on hybrid QGA-SVM[J]. Chinese Journal of Scientific Instrument, 2014, 35(2): 343-349.