Abstract:In order to investigate the accuracy and consistency of measurement values and issued measurement results for ophthalmometers by various laboratories, China National Institute of Metrology, as the pilot laboratory, organized and implemented the national Class A measurement comparison for radius verification ability of ophthalmometer from 2020 to 2021. A total of 33 laboratories participated in this measurement comparison, covering 15 provinces/municipalities directly under the Central Government, and the participating laboratories were highly representative. The comparison results were evaluated by the normalized error En value. With data processing and statistical analysis, the absolute value of En of 33 participated laboratories was between 0 and 0.85, all less than 1. The difference between the measurement results and reference values of each participated laboratory was within reasonable expectations. The results showed that the verification ability of radius of curvature in all participated laboratories met the technical requirements of measurement standards, and effectively ensured the accuracy, consistency and reliability of quantity values of ophthalmometers.
Zhang J Y, Hong B Y, Gao M L. Uncertainty evaluation for standard devices of ophthaomometers by using sample test method [J]. Measurement Technique, 2013(9): 47-50.
Zhang J Y, Wang L R, Chen Y. The development of standard devices for ophthalmometers [J]. Acta Metrologica Sinica, 2006, 27(1): 28-31.
Zhang J Y, Li J, Gao M L,et al . The development of axial standard device for calibration of ophthalmometers [J]. Acta Metrologica Sinica, 2017, 38(3): 276-279.
Zhang J Y, Liu W L, Gao M L, et al. Calibration and traceability method for human cornea parameters [J]. Acta Metrologica Sinica, 2020, 41(11): 1339-1343.
[11]
JJG 1011-2018 角膜曲率计 [S]. 2018.
[13]
JJG 892-2022 验光仪 [S]. 2022.
[15]
张以谟. 应用光学 [M]. 北京: 电子工业出版社, 2010.
Li Q Q, Wang Y, Zheng K J. Analysis of eye diopter, axial length and corneal curvature in children aged 6~18 years [J]. Journal of Preventive Medicine, 2020, 32(9): 917-919.
Li K R, Li Q L, Xu X Z, et al. Effect of axial length and corneal curvature radius and their ratio on refractive errors in children and adolescents [J]. Int Eye Sci, 2019, 19(10): 1667-1671.
Wang Y, K M, Wang W H, et al. Effects of axial length on corneal curvature, corneal astigmatism, anterior chamber depth and intraocular pressure in age-related cataract [J]. Recent Advances in Ophthalmology, 2017, 37(9): 879-882.
Chen Y, Hong B Y, Zhang J Y. Verification of eye refractometers with curvature measurement function of cornea or contact lens [J]. Measurement Technique, 2014(12): 65-67.
Zhang J Y, Liu W L, Gao M L, et al. Metrological calibration of ophthalmometers [C]//The 8th International Conference on BioMedcial Engineering and Informatics. Beijing, China, 2015:360-365.