1. College of Metrology and Measurement Engineering,China Jiliang University, Hangzhou, Zhejiang 310018, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:In order to enhance the measurement precision of the small angle, a method to measure small angle by two-arm non-common path non-localized interference of point light source is presented. In the method, the angle measurement system is built by establishing the theoretical model of small angle measurement, and the interference concentric rings are recorded by the CCD without lens and small pixs on the large target surface. The principle of the method is to fit a series of rings and circle center positioning method by multi-ring regression with both dark ring and bright ring. The method can measure the lateral shift of the center of the circle with high precision, and complete the comparison measurement experiment with the laser interference small angle measuring instrument. For the proposed angle system, the results show that the maximum fluctuation of the stability is -2.46″ in the x direction, and -1.92″ in the y direction within three hours; the repeatability had good agreement with the small angle laser interference within ±0.50″ which has a maximum deviation of -0.45″.
Wang Z C, Huang T, Kang Y T, et al. A quick-response and high-precision dynamic measurement method with application on small-angle measurement[J]. Measurement Science and Technology, 2022, 33: 045009.
Li X W, Li Q, Ji F, et al. Research on calibration method of initial incident angle of metrological laser ellipsometer[J]. Acta Metrologica Sinica, 2022, 43(5): 571-577.
Wang H L, Hu Y K, Zhang F. Design and Proficiency Test of the Measurement System for Small Angle Lasers[J]. Metrology & Measurement Technique, 2021, 48(11): 40-44.
Tan Y D, Xu X, Zhang S L. Precision measurement and application of laser interferometer[J]. Chinese Journal of Laser, 2021, 48(15): 1504001.
Song Z, Zhang J, Zhang H P, et al. Self-correction Method of Rolling Angle Measurement Error for Linear Guide Rail[J]. Acta Metrologica Sinica, 2022, 43(12): 1538-1543.
Tolansky S. LXIV. New contributions to interferometry. Part I. —New non-localized interference fringes[J]. The London, Edinburgh, and Dublin Philosophical Magazine and Journal of Science, 1943, 34(235): 555-565.
Bergman T G, Thompson J L. An Interference Method for Determining the Degree of Parallelism of (Laser) Surfaces[J]. Applied Optics, 1968, 7(5): 923-925.
[17]
Wang Y C, Shyu L H, Chang C P. Fabry-Pérot interferometer utilized for displacement measurement in a large measuring range[J]. Review of Scientific Instrumment, 2010, 81, 093102.
[18]
李星河. 激光干涉方法测量硅球直径研究[D]. 武汉: 华中科技大学, 2021.
Wang H B, Huang Y, Xue Z, et al. Measurement Uncertainty Evaluation for the Laser Small Angle Primary Standard[J]. Acta Metrologica Sinica, 2019, 40(6): 970-974.
Yang Z L, Fan B X, Xi Q, et al. Research on the Method of Building Space Rectangular Coordinate System of Photoelectric Autocollimator[J]. Acta Metrologica Sinica, 2018, 39(1): 12-14.
[8]
Gecheler R D, Schumann M, Just A, et al. A comparison of traceable spatial angle autocollimator calibrations performed by PTB and VTT MIKES[J]. Metrologia, 2022, 59: 024002.
Liu Y, Shen X Y, Zhou S N, et al. Micro-angle Measurement Method and Its Accuracy Evaluation Based on Fabry-Perot Etalon[J]. Acta Photonica Sinica, 2021, 50(7): 0712004.
Zhang G Y. Laser beam interferometry to measure parallel plain wedge angles and some extended applications[J]. Physics, 1974, 3(2): 119-122.
Zhou S N, Shen X Y, Li D S, et al. Self-calibration Correction Measurement of Micro-angle Based on F-P Etalon Multi-beam Interference Imaging[J]. Acta Photonica Sinica, 2022, 51(4): 0412007.