Development of Certified Reference Materials for Evaluation on Protection Performance of Biological Safety Cabinet
WANG Zi-quan1,LIU Si-yuan1,ZHEN Xiao-xiao2,HUNAG Wen-feng1,ZHANG Ling1,XU Qian3,4,SUI Zhi-wei1
1. Center for Advanced Measurement Science, National Institute of Metrology, Beijing 100029, China
2. Beijing Institute of Metrology, Beijing 100029, China
3. College of Life Sciences, Hunan Normal University, Changsha, Hunan 410081, China
4. Hunan Institute of Metrology and Test, Changsha, Hunan 410014, China
Abstract:The protection performance of biosafety cabinet not only affects the accuracy and reliability of experimental results, but also relates to the personal safety of laboratory personnel, therefore its protection performance needs to be evaluated and calibrated in time. Bacillus subtilis (B. subtilis) is an indicator of performance evaluation specified in the domestic and international standards of biosafety cabinet, thus development of B. subtilis reference material is necessary. Three reference materials were prepared by freeze-drying B. subtilis spores in the presence of a composite cryoprotectant. The result indicated that the reference materials of B. subtilis spores count bearing excellent homogeneity and could be stored stably for twenty-four months at 4℃. The certified values of three different concentrations of certified reference materials were (5.4±1.3)×108CFU/mL, (5.2±1.3)×106CFU/mL, (5.3±1.4)×104CFU/mL, respectively. The reference materials of B. subtilis spores count could mainly be used to evaluate the staff, product, and cross-contamination protection of the biosafety cabinet.
Liu J, Liu Y P, Zhang X H, et al. Study on Calibration Method of Biosafety Cabinet[J]. Metrology & Measurement Technique, 2020, 47(2):45-48.
Sui Z W, Wang B, Liu S Y, et al. Study on Measurement Method of Bacillus subtilis var. niger Spore[J]. Acta Metrologica Sinica, 2020, 41(9):1171-1176.
Huang X, Xu L L, Huang R S, et al. Research Advance in Controlling Plant Diseases by Bacillus subtilis[J]. Biotechnology Bulletin, 2010(1):24-29.
Liu S Y, Wang Z Q, Zhen X X, et al. Development of reference materials of Escherichia coli for productivity ratio evaluation of microbiological media[J]. China Measurement & Test, 2020, 46(10):48-53.
Chen S Z, Hu Y Y. Use of Bacillus subtilis in purification of slightly-polluted water[J]. Acta Scientiae Circumstantiae, 2011, 31(8):1594-1601.
[11]
Nakamura L K. Taxonomic relationship of black-pigmented Ba callus subtilis strains and aproposal for Bacillus atrophaeus sp. Nov[J]. Int J Syst Bacteriol, 1989, 39:295.
Xue L, Lin J, Sui Z W, et al. Preparation of Reference Material for Quantitative Detection of Escherichia Coli[J]. Acta Metrologica Sinica, 2015, 36(6):652-656.
Bi J J, Wen Z B, Zhao J J, et al. Study on Use of Serratia marcescens Test for Evaluation of Class Ⅱ Biosafety Cabinets[J]. Chinese Journal of Disinfection, 2007, 24(1):23-25.