为满足医药类生产企业中水质监测所使用的在线TOC分析仪在500μg/L浓度以下的原位校准需求,建立了一套由高精度微量陶瓷注射泵、磁力搅拌混匀单元和步进电机控制恒流注射器组成的在线水质TOC监测仪自动校准装置。通过PLC控制高精度微量陶瓷注射泵和大容量注射泵分别将高浓度TOC标准原液和纯水进行比例混匀,实现了对在线水质TOC监测仪500μg/L以下量程的自动校准,线性相关性0.9924。
Abstract
In order to meet the needs of in-situ calibration of online TOC analyzer used in water quality monitoring in pharmaceutical enterprises under 500μg/L concentration, a set of on-line total organic carbon (TOC) monitor in-situ calibration device is established, which is composed of high-precision micro ceramic injection pump, magnetic mixing unit and constant current syringe controlled by step motor. The high concentration TOC standard stock solution and pure water are proportioned by high precision micro ceramic injection pump and large capacity injection pump controlled by PLC. The automatic calibration of the measuring range below 500μg/L of the on-line total organic carbon (TOC) monitor is realized, with a linear correlation of 0.9924.
关键词
计量学 /
总有机碳 /
在线水质监测 /
原位校准
Key words
metrology /
total organic carbon /
on-line water quality monitoring;in-situ calibration
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
[1]Bayram A, nsoy H, Akinci G, et al. Variation of total organic carbon content along the stream Harsit, Eastern Black Sea Basin, Turkey [J]. Environ Monit Assess, 2011, 182: 85-95.
[2]国家药典委员会. 中华人民共和国药典[M]. 2010 版附录Ⅷ R. 北京: 中国医药科技出版社, 2010.
[3]罗明. 低含量(μg/L)级总有机碳分析仪校准方法的研究 [D]. 上海: 上海交通大学, 2016.
[4]JJG 821—2005总有机碳分析仪 [S].
[5]马康, 谷雪蔷, 黎朋. 总有机碳(TOC)分析技术及仪器的计量标准现状 [J]. 中国计量, 2011: 94-96.
[6]张爱亮, 刘悦, 陈启悦, 等. 一种气溶胶光度计质量浓度的校准方法 [J]. 计量学报, 2020, 41(6): 765-768.
Zhang A L, Liu Y, Cheng Q Y, et al. A Calibration Method for Mass Concentration of Aerosol Photometers [J]. Acta Metrologica Sinica, 2020, 41(6): 765-768.
[7]张遥奇, 任 昀, 李娅, 等. 一种纯水机内置在线监测仪表的现场校准方法 [J]. 计量学报, 2020, 41(1): 115-120.
Zhang Y Q, Ren J, Li Y, et al. A Calibration Method for Inner Installed on-line Monitoring Instrument of Pure Water Machine [J]. Acta Metrologica Sinica, 2020, 41(1): 115-120.
[8]孙悦. 总有机碳测定方法及最新应用进展 [J]. 天津药学. 2012, 24(1): 60-64.
[9]Lee B M, Veriansyah B, Kim S H. Total Organic Carbon Disappearance Kinetics for Supercritical Water Oxidation of Dimethyl Methylphospate Used as a Chemical Agent Simulant [J]. Korean J Chem Eng, 2005, 22(4): 579-584.
[10]Jung Y K, Kim M J, Kim Y J, et al. Limitation of UV-Vis Absorption Analysis for Determination of Aqueous Colloidal Fullerene (nC60) at High Ionic Strength [J]. KSCE Journal of Civil Engineering, 2013, 17(1): 51-59.
[11]邱灵佳, 黄国林, 苏玉, 等. 总有机碳测定方法研究进展 [J]. 广东化工, 2015, 42(9): 107-108.
Qiu L J, Huang G L, Su Y, et al. An Overview on Current Measurment Methods of Total Organic Carbon [J]. Guangdong chemical industry, 2015, 42(9): 107-108.
[12]徐惠忠, 徐滋秋, 付万年. 对总有机碳分析仪国产化问题的思考 [J]. 分析仪器, 2008(4): 51-56.
[13]安文昊, 夏慧琳, 苏永兴. 输液泵注射泵检测技术 [J]. 中国医疗设备, 2010, 25(1): 5-6.
An W H, Xia H L, Su Y X. Infusion Pump and Injection Pump Testing Technology [J]. China Medical Devices, 2010, 25(1): 5-6.
[14]张飚瑞, 杨威, 罗招平. 临床输液泵和注射泵的流速检测方法探讨 [J]. 中国医疗设备, 2011, 26(5): 87-88, 166.
Zhang B R, Yang W, Luo Z P. Discussion on the Test Methods for the Flow Rate of Syringe Pump and Infusion Pump [J]. China Medical Devices, 2011, 26(5): 87-88, 166.
[15]孙磊, 孙 斌, 赵玉晓. 双通道医用输液泵自动校准系统开发 [J]. 计量学报, 2019, 40(6): 1064-1070.
Sun L, Sun B, Zhao Y X. Research and Development of Dual-channel Medical Syringe Pump Automatic Calibration System [J]. Acta Metrologica Sinica, 2019, 40(6): 1064-1070.
[16]Mettler Toledo. 4000TOC传感器操作说明书[Z].
基金
江苏省市场监督管理局科研项目(KJ175920)