1. College of Metrology & Measurement Engineering, China Jiliang University, Zhejiang, Hangzhou 310018, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:A volume calibration method of spirometer syringes (hereinafter referred to as syringes) based on hydro-weighting method and combined with negative pressure principle is presented, and a calibration device for the volume of syringes is designed. During calibration, the medium (pure water) is pumped into the weighing cylinder by the negative pressure formed in syringes when the piston of syringes is pulled out. From the collected data such as ambient atmospheric pressure, medium mass and temperature, the capacity of syringes is calculated at 20°C. The volume measurement model is developed according to the volume measurement principle of syringes. The measured volume of a syringe is 3.00196L, and the uncertainty of measurement result is 9.6×10-4 (k=2). The research shows that the device meets the requirements of syringes capacity calibration and can realize rapid calibration.
Madsen F. Validation of spirometer calibration syringes[J]. Scand J of Clin Lab Invest, 2012, 72(8): 608-613.
[1]
Graham B L, Steenbruggen I, Miller M R, et al. Standardization of Spirometry 2019 Update. An Official American Thoracic Society and European Respiratory Society Technical Statement[J]. American Journal of Respiratory and Critical Care Medicine, 2019, 200 (8): e70-e88.
Zheng X Z. Several problems that should be noted in the pulmonary function test[J]. Journal of Clinical Pulmonary Medicine, 2011,16(10):1619.
Wu Z P, Huang R B, Zheng J P, et al. Development and application value assessment of a method for measuring the quality of volume calibration syringes[J]. Chinese General Practice, 2022, 25(2):149-152.
[11]
Wu Z P, Peng Y Y, Lin K Q, et al. Quality inspection and result analysis of the spirometer calibration cylinder[J]. BMC Pulm Med, 2022, 22: 218.
[17]
Batista E , Paton R. The Selection of Water Property Formulae for Volume and Flow Calibration [J]. Metrologia, 2007, 44,453-463.
Zhong J D, Sun B, Zhao, Y X, et al. Calibration Method for Optical Measurement of Ultra-Micro Liquid Volume[J]. Acta Metrologica Sinica, 2020, 41(8): 960-964.
Guo L G, Liu Z Y, Wang J T, et al. Study on Uncertainty of Volume Measerment by Hydrostatic Weighing Method and Production of Calibration Medium[J]. Acta Metrologica Sinica, 2013,34(6):534-540.
[22]
JJF1059-2012 测量不确定度评定与表示指南[S]. 2012.
Chen L F. Evaluating the Sampling Contribution to the Measurement Uncertainty[J]. Acta Metrologica Sinica, 2020, 41(7): 891-896.
[13]
Picard A, Davis R S, Glaser M, et al. Revised formula for the density of moist air (CIPM-2007) [J]. Metrologia, 2008, 45(2):149-155.
Tanaka M, Girard G, Davis R, et al. Recommended table for the density of water between 0℃ and 400℃ based on recent experimental reports [J]. Metrologia, 2001, 38:301-309.
[18]
EURAMET cg-19 Guidelines on the Determination of Uncertainty in Gravimetric Volume Calibration[S].
[21]
EURAMET cg-21(2013) Guidelines on the Calibration of Standard Capacity Measures Using the Volumetric Method[S].