Abstract:The Monte-Carlo (MC) model of the 90Sr/90Y source used in the BSS2 β-ray secondary standard and the PTW 23392 extrapolation chamber was constructed. The depth-dose curves produced by the 90Sr/90Y source at the calibrated distance were calculated using BEAMnrc and compared with the experimental results. According to the difference between the simulation results and the experimental results, the MC model of the 90Sr/90Y source was revised. In addition, the extrapolation curve of the extrapolation chamber, the tissue transmission factor and the tissue absorbed dose rate at the calibration position were simulated. Under the condition of 30cm away from the source and using the beam flattening filter, the calculated extrapolation curves are in good agreement with the experimental results. The differences between the simulated results of the transmission factor and the absorbed dose rate in tissue and the calibration certificate values are within 1.43%and 2.11%, respectively. The mentioned study constructed a more accurate MC model of the 90Sr/90Y source and extrapolation chamber. The results can provide a reference for the calculation of the β fluence spectrum, dose conversion factor and correction factors of the extrapolation chamber in β reference radiation field.
滕忠斌,宋明哲,王红玉,魏可新,刘蕴韬. 90Sr/90Y放射源及外推电离室蒙特卡罗模型建立与实验验证[J]. 计量学报, 2022, 43(11): 1512-1517.
TENG Zhong-bin,SONG Ming-zhe,WANG Hong-yu,WEI Ke-xin,LIU Yun-tao. The Monte Carlo Modeling and Experimental Verification of the 90Sr/90Y Source and the Extrapolation Chamber. Acta Metrologica Sinica, 2022, 43(11): 1512-1517.
[1]Hoedlmoser H, Greiter M, Bandalo V, et al. New eye lens dosemeters for integration in radiation protection glasses[J]. Radiation Measurements, 2019, 125:106-115.
[2]International Organization for Standardization. ISO 6980-1 Nuclear energy-Reference beta-particle radiation—Part 1:Methods of production[S]. 2019.
[3]International Organization for Standardization. ISO 6980-2 Nuclear energy-Reference beta-particle radiation—Part 2:Calibration fundamentals related to basic quantities characterizing the radiation field[S]. 2019.
[4]International Organization for Standardization. ISO 6980-3 Nuclear energy-Reference beta-particle radiation—Part 3:Calibration of area and personal dosemeters and the determination of their response as a function of beta radiation energy and angle of incidence[S]. 2019.
[5]Ambrosi P, Buchholz G, Helmstadter K. The PTB Beta Secondary Standard BSS 2 for radiation protection[J]. Journal of Instrumentation, 2007, 2(11):11002-11032.
[6]PTW-FREIBURG. Instruction Manual-Ionizing Radiation Detectors[EB/OL]. https://www.ptwdosimetry.com/en/.
[7]吴琦, 李德红, 黄建微,等. 基于90Sr源β射线的外推电离室入射窗修正因子[J]. 原子能科学技术, 2019, 53(5):920-925.
Wu Q, Li D H, Huang J W, et al. Extrapolation Chamber Entrance Window Correction Factor on 90Sr β Ray Radiation Field[J]. Atomic Energy Science and Technology, 2019, 53(5):920-925.
[8]王继, 吴钦宏,赵亮,等. 高剂量率近距离192Ir治疗源的井型电离室仪器的校准[J]. 计量学报, 2021, 42(9):1232-1236.
Wang J, Wu Q H, Zhao L, et al. Calibration of Well Type Ionization Chamber Instruments with High-dose-rate After-loading 192Ir Source[J]. Acta Metrologica Sinica, 2021, 42(9):1232-1236.
[9]吕雅竹, 赵瑞,李德红,等. X射线剂量当量仪校准因子不确定度评估[J]. 计量学报, 2020, 41(9):1133-1137.
Lü Y Z, Zhao R, Li D H, et al. Uncertainty Evaluation of Calibration Factor for X-ray Dose Equivalent Instrument[J]. Acta Metrologica Sinica, 2020, 41(9):1133-1137.
[10]张璇, 黄建微,李德红,等. β射线剂量仪校准因子不确定度分析[J]. 计量科学与技术,2022, 66(2):21-24.
Zhang X, Huang J W, Li D H, et al. Uncertainty Evaluation for the Calibration Factor of β-ray Dosemeters[J]. Metrology Science and Technology, 2022, 66(2):21-24.
[11]Bakshi A K, Vandana S, Selvam T P, et al. Measurement of the output of ISO recommended beta sources with an extrapolation chamber[J]. Radiation Measurements, 2013, 53-54:50-55.
[12]Faria F P, Reynaldo S, Fonseca T C F, et al. Monte Carlo simulation applied to the characterization of an extrapolation chamber for beta radiation dosimetry[J]. Radiation Physics & Chemistry, 2015, 116:226-230.
[13]Behrens R. Simulation of the radiation fields of the Beta Secondary Standard BSS 2[J]. Journal of Instrumentation, 2013, 8(2):2019-2042.
[14]Brunzendorf J. Depth-dose curves of the beta reference fields 147Pm, 85Kr and 90Sr/90Y produced by the beta secondary standard BSS2[J]. Radiation protection dosimetry, 2012, 151(2):211-217.
[15]Eckert & Ziegler Strahlen-und Medizintechnik AG. VZ-2104-001[EB/OL]. https://www.ezag.com.
[16]Rogers D W O, Walters B, Kawrakow I. BEAMnrc Users Manual, NRCC Report PIRS-0509[R]. Ottawa, National Research Council of Canada, 2000.
[17]International Commission on Radiological Protection. Nuclear Decay Data for Dosimetric Calculations[R]. ICRP Publication 107, 2007.
[18]Brunzendorf J. Determination of depth-dose curves in beta dosimetry[J]. Radiation protection dosimetry, 2012, 151(2):203-210.