高精度金-铂热电偶的制作和热电稳定性研究

郑玮,汤磊,向明东

计量学报 ›› 2022, Vol. 43 ›› Issue (2) : 191-195.

PDF(1281 KB)
PDF(1281 KB)
计量学报 ›› 2022, Vol. 43 ›› Issue (2) : 191-195. DOI: 10.3969/j.issn.1000-1158.2022.02.08
热学计量

高精度金-铂热电偶的制作和热电稳定性研究

  • 郑玮,汤磊,向明东
作者信息 +

The Research on Construction the High Precision Au/Pt Thermocouple and its Stability of Thermoelectric Character

  • ZHENG Wei,TANG Lei,XIANG Ming-dong
Author information +
文章历史 +

摘要

的关键。首先,对热电极丝材进行剪切、清洗、退火并焊接组装成热电偶;然后,再对制作成的热电偶进行退火实验,热电偶在其使用上限965℃附近累计退火时间700h,在银凝固点的稳定性达到±0.02℃。在制造热电偶过程中,采取了传统结构、小圈结构、U型结构3种不同的方式构建了热电偶测量端,实验结果表明:小圈结构制作的测量端其短期稳定性比起U型结构更好。

Abstract

The stability of thermo-electric character in long term application of Au/Pt thermocouple, which is fabricated by a pure golden wire and a platinum wire, is a key to high accurate performance to in temperature measurement.Firstly, the thermo-elements were cut, cleaned, annealed, welded. Then, the thermo-elements were assembled to the thermocouples and were tested. The stability of whole thermocouples was proved within ±0.02℃ at Ag freezing point calibration, after ageing time of thermocouples was summed up to about 700h at temperature up to 965℃.The thermocouple measurement junction was constructed in three different ways: traditional type, mini-coil type and U shape type, the experiment results showed that the short-term stability is better for mini-coil type compare to U shape type applied to thermocouple measurement junction connection.

关键词

计量学 / 金-铂热电偶;测量端 / 热电稳定性

Key words

metrology / Au/Pt thermocouple / measurement junction / thermoelectric stability

引用本文

导出引用
郑玮,汤磊,向明东. 高精度金-铂热电偶的制作和热电稳定性研究[J]. 计量学报. 2022, 43(2): 191-195 https://doi.org/10.3969/j.issn.1000-1158.2022.02.08
ZHENG Wei,TANG Lei,XIANG Ming-dong. The Research on Construction the High Precision Au/Pt Thermocouple and its Stability of Thermoelectric Character[J]. Acta Metrologica Sinica. 2022, 43(2): 191-195 https://doi.org/10.3969/j.issn.1000-1158.2022.02.08
中图分类号: TB942   

参考文献

[1]Quinn J T. Temperature[M]. New York & Lon-don: Academic Press, 1983:254-256.
[2]Bentley Robin E. Thermoelectric changes in Au and Pt: metals used in elemental thermocouples [J]. Measure-ment Science and Technology, 2001, 12(5): 627.
[3]Kim Y G, Gam K S, Rhee C. 1000h stability test of the Au/Pd thermocouple at 1000 ℃ [J]. Metrologia, 1996, 33(4): 337.
[4]Edler F, Arifovic N, Atance G, et al. Thermoelectric properties of currently available Au/Pt thermocouples related to the valid reference function [J]. International Journal of Metrology & Quality Engineering, 2015.
[5]凌善康. 国际温标补充资料 [M]. 北京: 中国计量出版社, 1992.
[6]Gotoh M, Hill K. D, Murdock E. G. A Gold/Platinum Thermocouple Reference Table [J]. Review of Scientific Instrument, 1991, 62(11): 2778 – 91.
[7]JJG 542—1997金-铂热电偶检定规程[S]. 1997.
[8]Kim Y G, Gam K S, Kang K H. Thermoelectric prop-erties of the Au/Pt thermocouple [J]. Review of Scientific Instruments, 1998, 69: 3577.
[9]Zheng W, Cheng W X. Researching the Platinum versus Palladium Thermocouple in the Range 419.527℃ to 1084.62 ℃ [C]// Proceedings of the AIP Conference Proceedings, New York, USA, 2003.
[10]Ferdouse J, Mark B. Stability Studies of a New Design Au/Pt Thermocouple Without a Strain Relieving Coil [J]. International Journal of Thermophysics, 2007, 28(6): 1822-31.
[11]Kim Y G, Yang I, Sool Gam K. Measurement of the Inhomogeneity of Pt/Pd Thermocouples using an Immer-sion-into-Liquid Method [J]. Instrumentation Science and Technology, 2008, 36(3): 257-66.
[12]向明东, 郑玮, 陈伟昕. 固定点分度装置校准铂铑10-铂热电偶不确定度评定 [J]. 现代测量与实验室管理, 2003, 11(2): 33-36.
 Xiang M D, Zheng W, Cheng W X. The Uncertailty evaluation of Fixed Point Instrument for PtRh10-Pt ther-mocouple calibration [J]. Advanced Measurement and Laboratory Management 2003, 11(2): 33-36.
[13]郑玮, 向明东. 铂铑10-铂热电偶在1100~1300℃外推计算公式的实验验证 [J]. 计量学报, 2007, 28(z1):69-72.
Zheng W,Xiang M D. Verification of Calculation Method of PtRh10-Pt Thermocouple in the Range of 1100℃ to 1300℃ by Experiment Techniques [J]. Acta Metro-logica Sinica, 2007, 28(z1):69-72.
[14]郑玮, 杨振华, 向明东. 在液槽中测试热电偶不均匀性方法探索 [J]. 计量学报, 2013, 34(4): 356-359.
Zheng W,Yang Z H, Xiang M D. Study of Inhomo-geneity Test on Thermocouple in Liquid Bath [J]. Acta Metrooogica Sinica, 2013, 34(4):356-359.
[15]向明东, 郑玮, 陈伟昕. 标准铂铑10—铂热电偶检定中的常见问题及解决方法 [J]. 计量技术, 2003 (3): 42-46.
[16]Gotoh M. Thermoelectric Scanning Study of Pt/Pd and Au/Pt Thermocouples up to 960 ℃ with a Pressure Controlled Sodium Heat-Pipe [J]. American Institute of Physics, 2003,684(1):481-484.
[17]郑玮, 向明东, 陈伟昕. 热电偶精密测温技术的发展方向 [J]. 现代测量与实验室管理, 2002 (4): 7-9.
[18 ]郑玮,汤磊. 标准铂铑10-铂热电偶热电势约束公式探讨 [J ] 计量学报, 2020, 41(2): 175-178.
Zheng W, Tang L. The Research on the Constraint Formulas of EMFs as a Standard Pt-10% Rh/Pt Therm-ocouple [J]. Acta Metrologica Sinica, 2020, 41(2):175-178.

基金

国家质量基础的共性研究与应用项目(2017YFF0205903)

PDF(1281 KB)

Accesses

Citation

Detail

段落导航
相关文章

/