Establishment of Reference Radiation Field for 40~125kV Pulsed X-ray
ZHANG Xiao-le1,2,HUANG Jian-wei2,LI De-hong2,CHENG Jian-bo2, YANG Yang2,CAO Lei2,3,ZHAO Rui2
1. Lanzhou University, Lanzhou, Gansu 730000, China
2. National Institute of Metrology, Beijing 100029, China
3. Beijing Normal University, Beijing 100011, China
Abstract:To solve the problem of timing correction of pulsed X-ray dose monitoring instruments for X-ray diagnosis and security inspection, 40~125kV pulsed X-ray reference radiation field that simulates the above-mentioned radiation conditions has been established. The reference radiation field have both conventional radiation field characteristics and pulse radiation field characteristics. A high-voltage generator based on grid control technology can control the X ray tube to generate X-rays with different pulse characteristics. To measure the pulse characteristics of the generated X-rays (rising and falling time, exposure time, etc.), the pulse time module has been used, and the results show that the reference radiation field can be used for time response correction and dose traceability of active dose monitoring equipment.
[1]Hupe O, Ankerhold U. Determination of ambient and personal dose equivalent for personnel and cargo security screening[J]. Radiation Protection Protection, 2016, 121(4): 274-278
[2]Ambrosi P, Borowski M, Iwatschenko M. Considerations concerning the use of counting active personal dosemeters in pulsed fields of ionizing radiation[J]. Radiation Protection Dosimetry, 2010, 139(4): 483-493.
[3]高飞, 陈东风, 肖雪夫, 等. 主动式电离辐射剂量仪的脉冲X射线响应特性研究[J]. 辐射防护, 2019, 39(6): 469-474.
Gao F, Chen D F, Xiao X F, et al. Study on pulsed X ray response characteristics of active ionization radiation dosimeters[J]. Radiation Protection, 2019, 39(6): 469-474.
[4]高飞, 徐阳, 刘蕴韬, 等. 脉冲X射线次级标准电离室研制及量传技术研究[J], 原子能科学技术, 2019, 53(11): 2259-2266
Gao F, Xu Y, Liu Y T, et al. Study on Pulse X Ray Secondard Ionization Chamber and Calibration Technology[J]. Atomic Energy science and technology, 2019, 53(11): 2259-2266
[5]Ambrosi P, Borowski M, Iwatschenko M. Considerations concerning the use of counting active personal dosemeters in pulsed fields of ionising radiation[J]. Radiation Protection Dosimetry, 2010, 193(4): 483-493
[6]Clairand J, Bordy M, Carinou E, et al. Use of active personal dosemeters in interventional radiology and cardiology: Tests in laboratory conditions and recommendations-ORAMED project[J]. Radiation Measurements, 2011, 46(11): 1252-1257
[7]周峰, 刘长军, 向元益, 等. 脉冲X射线辐射场监测问题探讨[J]. 中国辐射卫生, 2016, 25(3):261-264.
Zhou F, Liu C J, Xiang Y Y, et al. Discussion on Monitoring of Pulsed X-ray Radiation Field[J]. Chin J Radiol Health, 2016, 25(3): 261-264.
[8]乔凤普, 景敏卿, 刘恒, 等. 脉冲X、γ辐射场发生装置的研究[J]. 核电子学与探测技术, 2011, 31(9): 998-1002.
Qiao F P, Jing M Q, Liu H, et al. Research on Pulsed X-ray and γ-ray Radiation Field Generator[J]. Nuclear Electronics & Detection Technology, 2011, 31(9): 998-1002.
[9]张震, 侯长松, 梁婧, 等. 辐射检测仪器时间常数的测量与探讨[J]. 中华放射医学与防护杂志, 2017, 37(8): 609-613.
Zhang Z, Hou C S, Liang J, et al . Measurement and discussion of circuit time constantsfor radiation detectors[J]. Chin J Radiol med Prot, 2017, 37(8): 609-613
[10]周海伟, 杜国生. 两种电离室剂量仪响应时间的实验研究[J]. 中华放射医学与防护杂志, 2009, 29(6): 627-629.
Zhou H W, Du G S. Experiment and study of response time of two types of ion chamber survey meters[J]. Chin J Radiol Med Prot, 2009, 29(6). 627-629.
[11]欧向明. 常用辐射防护仪器电路时间常数的研究[J]. 中国医学装备, 2010, 7(2): 6-8
Ou X M. Research on circuits time constant of radiation protection meters[J]. China Med Equip, 2010, 7(2): 6-8
[12]Klammer J, Roth J, Hupe O. Novel reference radiation fields for pulsed photon radiation installed at PTB[J]. Radiation Protection Dosimetry, 2012, 151(3): 478-482
[13]IEC62743-2012 Radiation protection instrumention Electronic counting dosemeters for pulsed fields of ionizing radiation[S].
[14]宋飞, 赵瑞,丁卫撑,等. X射线环境辐射监测仪器校准方法的研究[J]. 计量学报, 2021, 42(9): 1237-1243.
Song F, Zhao R, Ding W C,et al. Research on Calibration Method of X-ray Environmental Radiation Monitoring Instrument[J]. Acta Metrologica Sinica, 2021, 42(9): 1237-1243.
[15]ISO 18090-1: 2015 Radiological protection characteristics of reference pulsed radiation-Part 1: Photon radiation[S].
[16]葛双, 吴金杰, 邱小平, 等. 工业钼靶X射线光机能谱的蒙特卡罗模拟[J]. 计量学报, 2015, 36 (6): 647-651.
Ge S, Wu J J, Qiu X P, et al. Monte Carlo Simulation for X-ray Spectrum On Industrial Molybdenum Target Units[J]. Acta Metrologica Sinica, 2015, 36(6): 647-651.
[17]ISO 4037-1: 1996 X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy—Part 1: Radiation characteristics and production methods[S].