2025年01月25日 星期六 首页   |    期刊介绍   |    编 委 会   |    投稿指南   |    期刊订阅   |    统合信息   |    联系我们
计量学报  2021, Vol. 42 Issue (6): 828-832    DOI: 10.3969/j.issn.1000-1158.2021.06.22
  电离辐射、标准物质与生物计量 本期目录 | 过刊浏览 | 高级检索 |
基于福尔马肼溶液的CRP快速分析仪检测方法
张宜文1,赵旭2,王骏1,隗永豪1,马康3,吴红1,赵海波1
1. 北京市计量检测科学研究院,北京 100012
2. 北京市海淀区田村社区卫生服务中心检验科,北京 100143
3. 中国计量科学研究院,北京 100029
Method of CRP Rapid Analyzer Based on Formazine Polymer
ZHANG Yi-wen1,ZHAO Xu2, WANG Jun1, WEI Yong-hao1, MA Kang3, WU Hong1,ZHAO Hai-bo1
1. Beijing Institute of Metrology, Beijing 100012, China
2.Tiancun Community Health Service Centers,  Beijing 100143, China
3. National Institute of Metrology, Beijing 100029, China
全文: PDF (1096 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 C-反应蛋白(C-reactive protein,CRP)快速分析仪是基层医疗机构常用的快速检验仪器之一,目前尚未有相应的规程和规范对此类型仪器的量值溯源和可靠性做为依据。通过分析福尔马肼溶液和C-反应蛋白溶液分子尺寸和分子量分布,依据散射原理并参照JJG 880—2006浊度计的方法项目对该仪器进行测试,并对方法的合理性进行评估。实验选用常见的Orion厂商的Quikread101(速率散射法)CRP快速分析仪,表明在浊度范围为(675~1000)NTU的福尔马肼溶液系列溶液测试时,测定线性及重复性良好。通过对系列福尔马肼溶液与C反应蛋白系列溶液进行对比,找到对应的福尔马肼溶液的浊度值,为该类仪器量值溯源提供了一种参考方法。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
张宜文
赵旭
王骏
隗永豪
马康
吴红
赵海波
关键词 计量学C反应蛋白快速分析仪免疫浊度福尔马肼溶液量值溯源    
Abstract:C-reactive protein (CRP) rapid analyzer is one of the commonly used rapid testing instruments in primary medical institutions. At present, there are no relevant regulations and specifications for traceability and reliability of this type of instrument. By analyzing the molecular size and molecular weight distribution of formazine solution and C-reactive protein solution, the instrument was tested on the basis of scattering principle and referring to the method items of JJG 880—2006 turbidimeter, and the rationality of the method was evaluated. Quikread101 (Rate Scattering) CRP rapid analyzer from Orion manufacturer was used in the experiment. The results showed that the linearity and repeatability of the determination were good when the formazine solution series was tested in turbidity range of (675~1 000) NTU. By comparing a series of formazine solutions with the series of C-reactive protein solutions, the corresponding turbidity values of formazine solutions were found, which provided a reference method for tracing the values of these instruments.
Key wordsmetrology;C-reactive protein rapid analyzer;immune turbidity;formazine solution    traceability
收稿日期: 2019-06-12      发布日期: 2021-06-23
PACS:  TB99  
基金资助:国家质量基础专项(2018YFF0212800,2018YFF0212801,2019YFF0216700);北京市科委计划(Z181100009518012)
通讯作者: 赵海波(1974-),北京人,北京市计量检测科学研究院高级工程师,研究方向为计量检测。Email:zhaohb@bjjl.cn     E-mail: zhangyw@bjjl.cn
作者简介: 张宜文(1985-),浙江丽水人,北京市计量检测科学研究院高级工程师,研究方向为计量检测。Email: zhangyw@bjjl.cn
引用本文:   
张宜文,赵旭,王骏,隗永豪,马康,吴红,赵海波. 基于福尔马肼溶液的CRP快速分析仪检测方法[J]. 计量学报, 2021, 42(6): 828-832.
ZHANG Yi-wen, ZHAO Xu, WANG Jun, WEI Yong-hao, MA Kang,WU Hong, ZHAO Hai-bo. Method of CRP Rapid Analyzer Based on Formazine Polymer. Acta Metrologica Sinica, 2021, 42(6): 828-832.
链接本文:  
http://jlxb.china-csm.org:81/Jwk_jlxb/CN/10.3969/j.issn.1000-1158.2021.06.22     或     http://jlxb.china-csm.org:81/Jwk_jlxb/CN/Y2021/V42/I6/828
京ICP备:14006989号-1
版权所有 © 《计量学报》编辑部
地址:北三环东路18号(北京1413信箱)  邮编:100029 电话:(010)64271480
本系统由北京玛格泰克科技发展有限公司设计开发  技术支持:support@magtech.com.cn