Abstract:Aiming at the problem that the all-digital driving technology flow tube is difficult to start vibration, five kinds of starting methods such as analog to digital start-up, random wave start-up, sine wave start-up, sweep frequency square wave start-up, and positive-negative step alternate start-up are reviewed. Aiming at the problem that the amplitude, frequency and phase of driving signal are difficult to adjust, the solution is proposed. For the implementation platform of all-digital drive technology, two solutions are summarized: a single-core all-digital drive based on DSP and a dual-core all-digital drive based on DSP+FPGA. Among them, the single-core all-digital driving based on DSP has low cost and is easy to promote, and the dual-core all-digital driving based on DSP+FPGA has higher precision.
[1]Anklin M, Drahm W, Rieder A. Coriolis mass flowmeters: Overview of the current state of the art and latest research [J]. Flow Measurement & Instrumentation, 2006, 17(6): 317-323.
[2]梁平, 岳峥, 刘晨曦. 数字驱动技术在科氏质量流量计中的应用 [J]. 自动化技术与应用, 2018, 37(1): 122-124.
Liang P, Yue Z, Liu C X. Application of Digital Drive Technology for Coriolis Mass Flowmeter [J]. Techniques of Automation and Application, 2018, 37(1): 122-124.
[3]李苗, 徐科军, 朱永强, 等. 科氏质量流量计的3种驱动方法研究 [J]. 计量学报, 2011, 32(1): 36-39.
Li M, Xu K J, Zhu Y Q, et al. Studies on three driving methods of Coriolis mass flowmeter [J]. Acta Metrologica Sinica, 2011, 32(1): 36-39.
[4]Seeger M. Coriolis flow measurement in two phase flow [J]. Computing & Control Engineering Journal, 2005, 16(3): 10-16.
[5]Basse N T. Coriolis flowmeter damping for two-phase flow due to decoupling [J]. Flow Measurement & Instrumentation, 2016, 52.
[6]Kunze J W, Storm R, Wang T. Coriolis Mass Flow Measurement with Entrained Gas[C]//Sensors and Measuring Systems 2014,17 ITG/GMA Symposium. Nuremberg, Germany,2014.
[7]Charreton C, Béguin C, Ross A, et al. Two-phase damping for internal flow: Physical mechanism and effect of excitation parameters [J]. Journal of Fluids & Structures, 2015, 56: 56-74.
[8]徐科军, 刘文, 乐静, 等. 一种基于FPGA的科氏质量流量计数字驱动系统:CN107167196A[P]. 2017.
[9]Liu R P, Fuent M J, Henry M P, et al. A neural network to correct mass flow errors caused by two-phase flow in a digital Coriolis mass flowmeter [J]. Flow Measurement & Instrumentation, 2001, 12(1): 53-63.
[10]Hou Q L, Xu K J, Fang M, et al. Gas-liquid two-phase flow correction method for digital CMF [J]. IEEE Transactions on Instrumentation and Measurement, 2014, 63(10): 2396-2404.
[11]陶波波, 侯其立, 石岩, 等. 科氏质量流量计测量含气液体流量的方法与实现 [J]. 仪器仪表学报, 2014(8): 1796-1802.
Tao B B, Hou Q L, Shi Y, et al. Method and implementation of measuring the flow of liquid mixed with gas for Coriolis mass flowmeter [J]. Chinese Journal of Scientific Instrument, 2014, 35(8): 1796-1802.
[11]马亚朋,石础,罗宇, 等. 科氏流量计气-液两相流的数值模拟[J]. 计量学报, 2018, 39(5): 663-667.
Ma Y P, Shi C, Luo Y, et al. Numerical Simulation of Gas-liquid Two-phase Flow in Coriolis Flowmeter[J]. Acta Metrologica Sinica, 2018, 39(5): 663-667.
[12]刘文, 徐科军, 方正余, 等. 用于振幅控制的科氏流量管模型的实验识别 [J]. 电子测量与仪器学报, 2018, 32(6): 39-45.
Liu W, Xu K J, Fang Z Y, et al. Experimental Identification of a Coriolis Flow Tube Model for Amplitude Control [J]. Journal of Electronic Measurement & Instrument, 2018, 32(6): 39-45.
[13]熊文军, 徐科军, 方敏, 等. 科氏流量计一次仪表与变送器匹配方法研究 [J]. 电子测量与仪器学报, 2012, 26(6): 521-528.
Xiong W J, Xu K J, Fang M, et al. Study on matching method of primary meter and transmitter in Coriolis mass flowmeter [J]. Journal of Electronic Measurement & Instrument, 2012, 26(6): 521-528.
[14]HIDEKISHIMADA. Coriolis flowmeter:USPatent 2010/0268484A1[P]. 2010.
[15]Henry M P, Zamora M E. Startup and operational techniques for a digital flowmeter: US 8483979 B2[P]. 2013.
[16]徐科军, 侯其立, 方敏, 等. 一种科氏质量流量计的数字驱动跟踪方法和系统:CN201110331814. X[P]. 2011.
[17]李京华, 黄熠彬. 一种基于全数字驱动方式的科氏质量流量计启振方法:CN107167195A[P]. 2017.
[18]李苗, 徐科军, 侯其立, 等. 数字科氏质量流量计正负阶跃交替激励启振方法 [J]. 仪器仪表学报, 2010, 31(1): 172-177.
Li M, Xu K J, Hou Q L, et al. Startup method of digital Coriolis mass flowmeter using alternating exciting of positive-negative step signal [J]. Chinese Journal of Scientific Instrument, 2010, 31(1): 172-177.
[19]李祥刚, 徐科军. 科氏质量流量管非线性幅值控制方法研究 [J]. 电子测量与仪器学报, 2009, 23(6): 82-86.
Li X G, Xu K J. Research on non-linear amplitude control method of Coriolis mass flow-tube [J]. Journal of Electronic Measurement & Instrument, 2009, 23(6): 82-86.
[20]李祥刚, 徐科军, 朱永强. 非线性幅值控制算法中2种PI控制器的比较 [J]. 合肥工业大学学报(自然科学版), 2009, 32(11): 1665-1668+1690.
Li X G, Xu K J, Zhu Y Q. Comparison between two types of PI controllers in non-linear amplitude control algorithm [J]. Journal of Hefei University of Techology, 2009, 32(11): 1665-1668+1690.
[21]陶波波, 徐科军, 侯其立, 等. 变传感器设定值的科氏质量流量管控制方法 [J]. 仪器仪表学报, 2015, 36(3): 712-720.
Tao B B, Xu K J, Hou Q L, et al. Variable sensor-setting-value based control method for Coriolis Mass Flowtube flowmeter [J]. Chinese Journal of Scientific Instrument, 2015, 36(3): 712-720.
[22]刘文, 徐科军, 乐静, 等. 科氏流量计幅值控制中两类关键参数的确定 [J]. 电子测量与仪器学报, 2018, 32(10): 183-189.
Liu W, Xu K J, Yue J, et al. Determination of two types of key parameters in coriolis flowmeter amplitude control [J]. Journal of Electronic Measurement & Instrument, 2018, 32(10): 183-189.
[23]张建国, 徐科军, 方正余, 等. 数字信号处理技术在科氏质量流量计中的应用 [J]. 仪器仪表学报, 2017, 38(9): 2087-2102.
Zhang J G, Xu K J, Fang Z Y, et al. Applications of digital signal processing technology in Coriolis mass flowmeter [J]. Chinese Journal of Scientific Instrument, 2017, 38(9): 2087-2102.
[24]徐科军, 倪伟. 一种科里奥利质量流量计的信号处理方法 [J]. 计量学报, 2001, 22(4): 254-287.
Xu K J, Ni W. A new signal processing method for Coriolis mass flowmeter [J]. Acta Metrologica Sinica, 2001, 22(4): 254-257.
[25]Jacobsen E, Lyons R. The sliding DFT [J]. Signal Processing Magazine IEEE, 2003, 20(2): 74-80.
[26]刘翠, 徐科军, 侯其立, 等. 适用于频繁启停流量测量的科氏质量流量计信号处理方法 [J]. 计量学报, 2014, 35(3): 242-247.
Liu C, Xu K J, Hou Q L, et al. Signal processing method for Coriolis mass flowmeters for frequent start-stop flow measurement [J]. Acta Metrologica Sinica, 2014, 35(3); 242-247.
[27]张建国, 徐科军, 董帅, 等. 基于希尔伯特变换的科氏质量流量计信号处理方法研究与实现 [J]. 计量学报, 2017, 38(3): 309-314.
Zhang J G, Xu K J, Dong S, et al. Study and Implementation of Signal Processing Method for Coriolis Mass Flowmeter Based on Hilbert Transform [J]. Acta Metrologica Sinica, 2017, 38(3), 309-314.
[28]刘维来, 赵璐, 王克逸, 等. 基于希尔伯特变换的科氏流量计信号处理 [J]. 计量学报, 2013, 34(5): 446-451.
Liu W L, Zhao L, Wang K Y, et al. Signal processing for Coriolis mass flowmeter based on Hilbert transform [J]. Acta Metrologica Sinica, 2013, 34(5): 446-451.
[29]侯其立, 徐科军, 李叶, 等. 用于微弯型科氏质量流量计的数字变送器研制 [J]. 电子测量与仪器学报, 2011, 25(6): 540-545.
Hou Q L, Xu K J, Li Y, et al. Development of digital transmitter for micro-bend type coriolis mass flowmeter [J]. Journal of Electronic Measurement and Instrument, 2011, 25(6): 540-545.
[30]郑德智, 樊尚春, 邢维巍. 科氏质量流量计相位差检测新方法 [J]. 仪器仪表学报, 2005,26(5): 441-443.
Zheng D Z, Fan S C, Xing W W. The novel method of phase difference detection in Coriolis Mass Flowmeter [J]. Chinese Journal of Scientific Instrument, 2005, 26(5): 441-443.
[31]杨俊, 武奇生, 孙宏琦. 基于相关法的相位差检测方法在科氏质量流量计中的应用研究 [J]. 传感器技术学报, 2007, 20(1): 138-145.
Yang J, Wu Q S, Sun H Q. Study on using correlation to detect phase difference in Coriolis Flowmeter [J]. Chinese Journal of Sensors and Actuators, 2007, 20(1): 138-145.
[32]涂亚庆, 沈廷鳌, 李明, 等. 基于多次互相关的非整周期信号相位差测量算法 [J]. 仪器仪表学报, 2014, 35(7): 1578-1585.
Tu Y Q, Shen T A, Li M, et al. Research on phase difference measurement algorithm for non-integer period sampling signal based on multi-layer correlation [J]. Chinese Journal of Scientific Instrument, 2014, 35(7): 1578-1585.
[33]侯其立, 徐科军, 方敏, 等. 科氏质量流量计数字驱动方法研究与实现 [J]. 计量学报, 2013, 34(6): 554-560.
Hou Q L, Xu K J, Fang M, et al. Study and implementation of digital drive method for Coriolis mass flow meter [J]. Acta Metrologica Sinica, 2013, 34(6): 554-560.
[34]Zamora M, Henry M P. An FPGA Implementation of a Digital Coriolis Mass Flow Metering Drive System [J]. IEEE Transactions on Industrial Electronics, 2008, 55(7): 2820-2831.
[35]Henry M P, Clarke D W, Archer N, et al. A self-validating digital Coriolis mass-flow meter: an overview [J]. Control Engineering Practice, 2000, 8(5): 487-506.
[36]田婧, 樊尚春, 郑德智. 基于FPGA的科氏质量流量计数字闭环系统设计 [J]. 仪表技术与传感器, 2009,(s1): 381-383.
Tian J, Fan S C, Zheng D Z. Design of Digital Closed Loop System of Coriolis Mass Flowmeter Based on FPGA [J]. Instrument Technique & Sensor, 2009,(s1): 381-383.
[37]陈宝欣, 涂亚庆, 杨辉跃, 等. 科氏流量计数字驱动系统设计 [J]. 后勤工程学院学报, 2016, 32(3): 86-91.
Chen B X, Tu Y Q, Yang H Y, et al. Method and Implementation of a Digital Coriolis Mass Flowmeter Drive System [J]. Journal of Logistical Engineering University, 2016, 32(3): 86-91.
[38]Hou Q L, Xu K J, Fang M, et al. Development of Coriolis mass flowmeter with digital drive and signal processing technology [J]. Isa Transactions, 2013, 52(5): 692.
[39]李苗. 科氏质量流量计数字驱动方法与两相流实验装置[D]. 合肥:合肥工业大学, 2011.