Abstract:The method of preparing high-accuracy standard radon source by using the absolute activity measuring device was studied, and the source of the value transfer with the uncertainty of activity expansion of 0.8% (k=2) was obtained. According to the basic principle of the calculation of the standard radon volume activity,calibrated the effective volume of the small dedicated radon chamber for value transfer, the expanded uncertainty of its effective volume is 0.46% (k=2). Calibration of the reference instrument (AlphaGUARD type standard measuring instrument for radon) used in general standard equipment (standard radon chamber), obtained the average volume activity response is 1.01, and the expanded uncertainty of the volume activity response is 3.2% (k=2).
闫钇帆,梁珺成,杨志杰,王仁波. 标准测氡仪的量值溯源方法研究[J]. 计量学报, 2020, 41(2): 250-256.
YAN Yi-fan,LIANG Jun-cheng,YANG Zhi-jie,WANG Ren-bo. Research on the Method of Traceability the Value of Standard Radon Monitor. Acta Metrologica Sinica, 2020, 41(2): 250-256.
[1]UNSCEAR. Source Effects and Risks of Ionizing Radiation[R]. New York: Unite Nations, 2000.
[2]Ting D S K .WHO Handbook on Indoor Radon: A Public Health Perspective[J]. International Journal of Environmental Studies, 2010, 67(1):100-102.
[3]吴凯华, 张相山, 王国龙, 等. 浅谈量值传递和量值溯源的实施[J]. 计量与测试技术, 2013, 40(1): 52-53.
Wu K H,Zhang X S,Wang G L,et al.Discussion Implementation on the Quantity Value Transfer and Measurement[J]. Traceability.Metrology & Measurement Technique,2013,40(1):52-53.
[4] Colle R, Hutchinson J M R, Unterweger M P.NIST primary radon-222 measurement system[J]. Journal of Research of the National Institute of Standards and Technology . 1990, 95(2):155-165.
[5]张曦, 邢雨, 肖雪夫. 氡活度绝对测量中氡源最佳冷凝半径的ANSYS模拟[J]. 原子能科学技术, 2016, 50(3): 565-570.
Zhang X,Xing Y,Xiao X F. Simulation of Best Condensation Radius of Radon Source Based on ANSYS in Absolute Measurement of Radon Activity[J]. Atomic Energy Science and Technology,2016,50(3):565-570.
[6]Busch I , Greupner H, Keyser U. Absolute measurement of the activity of 222Rn using a proportional counter[J]. Nuclear Instruments & Methods in Physics Research(section A), 2002, 481(1-3):330-338.
[7]张洋, 徐明,孙高峰, 等.氡室建标中几个技术问题的解决办法[J]. 计量学报, 2017, 38(2): 234-237.
Zhang Y,Xu M,Sun G F,et al.Research on the Progeny Volume Activty Measuring Method of 222Rn with Liquid Scintillation Counting[J].Acta Metrologica Sinica, 2017, 38(2): 234-237.
[8]王沥苑, 汤彬,梁珺成, 等. 氡绝对测量装置的小立体角计算方法研究[J]. 计量学报, 2016, 37(6A): 115-119.
Wang L Y,Tang B,Liang J C, et al.Small Solid Angle Calculation of Radon Absolute Measurement Apparatus[J].Acta Metrologica Sinica, 2016, 37(6A): 115-119.
[9]杨志杰, 李则书,侯胜利, 等.氡子体体积活度的液体闪烁计数方法研究[J]. 计量学报, 2015, 36(3): 328-332.
Yang Z J,Li Z S,Hou S L, et al.Research on the Progeny Volume Activty Measuring Method of 222Rn with Liquid Scintillation Counting[J].Acta Metrologica Sinica, 2016, 36(3): 328-332.
[10]Liang J C , Yang Z J, Wang L Y, et al. Development of the absolute standardization apparatus for radon-222 activity[J]. Applied Radiation and Isotopes, 2018, (134):358-362.
[11]Liang J C, Zheng P H, Yang Z J, et al.Development of calibration facility for radon and its progenies at NIM (China)[J]. Radiation Protection Dosimetry, 2015, 167(1-3):82-86.
[12]李泽西. 氡活度的小立体角绝对测量方法研究[D]. 衡阳:南华大学, 2017.
[13]刘翠红, 张磊, 卓维海, 等. AlphaGUARD测氡仪的~(220)Rn响应研究[J]. 辐射防护, 2010, 30(3): 135-140.
Liu C H,Zhang L,Zhuo W H,et al.Response of AlphaGUARD Radon Detector to 220Rn[J].Radiation Protection,2010,30(3):135-140.
[14]陆小军, 何林锋, 徐一鹤, 等. AlphaGUARD在主/被动模式下对~(222)Rn体积活度的响应[J]. 核电子学与探测技术, 2014, 34(10): 1192-1195.
Lu X J,He L F,Xu Y H,et al.The 222Rn Volume Radioactivity Response of AlphaGUARD Radon Monitor in Active/passive Mode[J].Nuclear Electronics & Detection Technology,2014,34(10):1192-1195.
[15]Wanke C , Szermerski B, Geworski L. Measurement of Rn-219 with Alphaguard radon monitors[J]. Applied Radiation and Isotopes, 2019,151:310-316.
[16]徐明, 孙高峰, 张洋, 等. 低气压地区闪烁室和PQ2000测氡仪的刻度校准研究[J]. 核电子学与探测技术, 2015, 35(3): 310-312.
Xu M,Sun G F,Zhang Y,et al.
The Calibration Study of Flash Room and PQ2000 Radon Detector in Low Airpressure Area[J]. Nuclear Electronics & Detection Technology,2015,35(3):310-312.