Abstract:To solve the problems of low efficiency, low detection accuracy, temperature deviation and data batch processing in the manual calibration process of medical infusion pump, an infusion pump automatic calibration scheme based on indirect flow method was proposed. The volume of the specific glass tube segment was calibrated through experiments to obtain the cumulative flow volume, and the accumulative time was obtained by measuring the time of filling the calibrated volume when the infusion pump was automatically calibrated. Thus a dual-channel medical infusion pump automatic calibration system was designed, which can perform error and repeatability analysis and uncertainty evaluation of the measurement results and give a calibration report. The verification experiment shows that the relative error of the automatic calibration system can reach less than 2%, meeting the requirements of 5% of the relevant specifications. The system can repeatedly measure the accuracy of the same infusion pump at different flow rates, which can also use dual channels to simultaneously perform flow calibration tests on multiple infusion pumps, as well as to calibrate the blocking alarm condition of the infusion pump.
孙磊,孙斌,赵玉晓. 双通道医用输液泵自动校准系统开发[J]. 计量学报, 2019, 40(6): 1064-1070.
SUN Lei,SUN Bin,ZHAO Yu-xiao. Research and Development of Dual-channel Medical Syringe Pump Automatic Calibration System. Acta Metrologica Sinica, 2019, 40(6): 1064-1070.
[1]Rigel Medical. Rigel Multi-Flo Key Benefits [EB/OL]. http://www.docin.com/p-1667429187.html, 2018-03-10.
[2]张平, 刘丽琼. 输液泵安全应用的全程管理[J]. 中国医疗设备, 2009, 24(7): 85-100.
Zhang P, Liu L Q. Complete Management of Safe Application for Infusion Pumps [J]. China Medical Devices, 2009, 7(24): 85-100.
[3]吴月明. 医用注射泵工作原理与质量控制[J]. 计量与测试技术, 2013, 40(6): 49-51.
Wu Y M. Working Principle and Quality Control of Hospital Syringe Pump [J]. Metrology & Measurement Technique, 2013, 40(6): 49-51.
[4]刘晓菲. 便携式智能输液器的研究[D]. 青岛:中国海洋大学, 2010: 22-25.
[5]张永胜, 刘彦军, 王金华. 利用Micro-PIV测量微管道内流量的研究[J]. 计量学报, 2011, 32(3): 209-212.
Zhang Y S, Liu Y J, Wang J H. The Flowrate Measurement in Microtube by Micro-PIV [J]. Acta Metrologica Sinica, 2011, 32(3): 209-212.
[6]王书涵. 浅谈医用输液泵的校准[J]. 计量与测试技术, 2012, 39(5): 47-50.
Wang S H. Discussion on the Calibration of Medical Infusion Pumps [J]. Metrology & Measurement Technique, 2012, 39(5): 47-50.
[7]安文昊, 夏慧琳, 苏永兴. 输液泵注射泵检测技术[J]. 中国医疗设备, 2010, 25(1): 5-6.
An W H, Xia H L, Su Y X. Infusion Pump and Injection Pump Testing Technology [J]. China Medical Devices, 2010, 25(1): 5-6.
[8]张飚瑞, 杨威, 罗招平. 临床输液泵和注射泵的流速检测方法探讨[J]. 中国医疗设备, 2011, 26(5): 87-88,166.
Zhang B R, Yang W, Luo Z P. Discussion on the Test Methods for the Flow Rate of Syringe Pump and Infusion Pump [J]. China Medical Devices, 2011, 26(5): 87-88,166.
[9]Vásquez J D H, Orlando A, Maurício F, et al. An alternative gravimetric measurement standard for calibration of liquid flow meters [J]. Flow Measurement and Instrumentation, 2017, 58: 87-96.
[10]Koech R. Water density formulations and their effect on gravimetric water meter calibration and measurement uncertainties [J]. Flow Measurement and Instrumentation, 2015, 45: 188-197.
[11]Bray A, Barbato G, Franceschini F. The effect of gravity acceleration on non-automatic weighing systems [J]. Measurement, 1995, 15 (4): 261-272.
[12]赵彩云. 基于LabVIEW的医用泵批量计量检测系统中压力和流量的检测[D]. 济南:山东中医药大学, 2014.
[13]沈昱明, 张进明. 对流量测量不确定度评估中若干名词术语的看法[J]. 计量学报, 2016, 37(1): 109-112.
Shen Y M, Zhang J M. Perspective on Some Terminology of Evaluation of Flow Measurement Uncertainty [J]. Acta Metrologica Sinica, 2016, 37(1): 109-112.
[14]张磊. 医用注射泵流量基本误差测量值的不确定度评定[J]. 计量与测试技术, 2015, 42 (7): 92-95.
Zhang L. Uncertainty Evaluation of Basic Error of Flow for Syringe Pumps [J]. Metrology & Measurement Technique, 2015, 42 (7): 92-95.
[15]李之昊,王金涛,刘翔,等. 基于Karl Fischer电位滴定原理的超微量液体容积精密测量方法[J]. 计量学报, 2017, 38(3): 296-299.
Li Z H, Wang J T, Liu X, et al. Non-Gravimetric Method For Ultra Micro Liquid Volume Metrology[J]. Acta Metrologica Sinica, 2017, 38(3): 296-299.
[16]李震. 注射泵(输液泵)流量测量结果不确定度分析[J].中国医疗器械信息, 2012, (9): 24-25, 61.
Li Z. Uncertainty Analysis of Injection(Infusion) Pump Flow Measurement Result [J]. China Medical Device Information, 2012, (9): 24-25, 61.