计量学报  2017, Vol. 38 Issue (6): 799-802    DOI: 10.3969/j.issn.1000-1158.2017.06.31
  电离辐射、标准物质与生物计量 本期目录 | 过刊浏览 | 高级检索 |
全自动酶免分析系统校准技术研究
康娟1,武利庆2,严振宇1,王军1,孙京昇1
1.北京市医疗器械检验所 医疗器械检验与安全性评价北京市重点实验室, 北京 101111
2.中国计量科学研究院 医学生物所, 北京 100029
Technical Research on Calibration of Fully Automated ELISA Analyzers
KANG Juan1,WU Li-qing2,YAN Zhen-yu1,WANG Jun1,SUN Jing-sheng1
1.Beijing Institute of Medical Device Testing & Beijing Key Lab of Medical Device Testing & Safety Evaluation, Beijing 101111, China
2. Division of Medical and Biological Measurement, National Institute of Metrology, Beijing 100029, China
全文: PDF (320 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 研究了一种用于评价全自动酶免分析系统测试性能的校准技术。该技术分别就加样模块的加样正确度和精密度,孵育模块的孵育温度,洗板模块的洗涤残留率和交叉污染,读板模块的吸光度示值稳定性、示值误差、重复性和通道差异进行测试和评价。选取3款不同品牌设备进行适用性验证,试验结果表明:所选系统最大加样偏差和变异系数分别为-7.4%和8.0%;孵育温度偏差在±0.5 ℃范围内;最大洗涤残留率和交叉污染分别为0.8%和0.1%;10 min内最大示值波动为0.003,在所测波长和吸光度下的最大示值误差为0.019,最大变异系数为0.2%,最大通道差异为0.005。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
康娟
武利庆
严振宇
王军
孙京昇
关键词 计量学酶免分析系统校准    
Abstract:A calibration technique has been established to evaluate the testing performance of fully automated ELISA analyzers. The technique tests the accuracy and precision of adding module, the temperature of incubation module, the washing residue rate and cross contamination of washing module, and the absorbance stability, accuracy, precision and channel difference of reading module separately around the constituent modules of the system. 3 fully automated ELISA analyzers from different manufactures were chosen to validate the technique applicability. Results showed that the maximum adding bias and CV were -7.4% and 8.0% respectively, the incubation temperature bias was within ±0.5 ℃, the maximum washing residue rate was 0.8% and maximum cross contamination was 0.1%, the maximum absorbance fluctuation was 0.003 within ten minutes, the maximum absorbance bias and CV were 0.019 and 0.2% respectively for tested wavelengths and absorbance,the maximum channel difference was 0.005. The calibration technique was testified to be scientific and practical, and expected to be used to evaluate the fully automated ELISA analyzers effectively.
Key wordsmetrology    ELISA analyzers    calibration
收稿日期: 2016-11-03      发布日期: 2017-09-27
PACS:  TB99  
基金资助:公益性行业科研专项(201310008-02)
通讯作者: 康娟     E-mail: kang_bimt@163.com
作者简介: 康娟(1978-),女,山西介休人,北京市医疗器械检验所高级工程师,主要从事体外诊断标准化方面的研究。 kang_bimt@163.com
引用本文:   
康娟,武利庆,严振宇,王军,孙京昇. 全自动酶免分析系统校准技术研究[J]. 计量学报, 2017, 38(6): 799-802.