Abstract:For the liquid volume measurement of volumetric glass gauge, a liquid level measurement method for micro liquid volume based on laser and machine vision technology was proposed. A volume table for correspondence relationship of liquid level and capacity in the standard glass gauge was established. The corresponding value of liquid volume can be obtained by measuring the liquid level. Measuring the liquid level of the standard glass gauge is the main point in the method. Liquid level measurement combined with computer and digital image processing technology, and used high-resolution camera and image processing software to extract feature element. High-precision volume measurement was realized, a comparative test with the static measure method was also designed, and validated the method. For the 100mL glass container, the measuring accuracy is better than 0.2%, the expanded uncertainty of the fitting algorithm is 0.137% (k=2), and the repeatability of measurement is 0.08%.
[1]左世全. 我国应将发展智能制造业提升到战略高度[J]. 中国科技投资, 2012, (31):47-50.
[2]李领录, 张珑. 专用玻璃量器检定数据处理自动化系统的设计和应用[J]. 计量技术, 2013, (8):76-80.
[3]佟林, 王金涛, 刘子勇, 等. 100 μL微量移液器国际关键比对的数学模型研究[J]. 计量学报, 2015, 36(2):154-157.
[4]王金涛, 陈超云, 李志昊, 等. 超微量液体容积双波长参比测量方法研究[J]. 计量学报, 2015, 36(3):256-259.
[5]Madec T, Amarouche S. Comparison of micropipet tescali bration by accredited laboratories[C]//CAFMET. 3rd International Metrology Conference 2010.Cairo,Egypt,2010:337-343.
[6]Stangegaard M, Hansen A J, Frslev T G, et al. A Simple method for validation and verification of pipettes mounted on automated liquid handlers[J]. Journal of Laboratory Automation, 2011,16(5):381-386.
[7]ISO 8655-7-2005 Piston-operated volumetric apparatus-Part 7:Non-gravimetric methods for the assessment of equipment performance[S].
[8]李之昊,王金涛,刘翔, 等. 基于Karl Fischer电位滴定原理的超微量液体容积精密测量方法[J]. 计量学报, 2017, 38(3): 296-299.
[9]刘小力. 机器人视觉伺服系统的图像处理和标定技术研究[D]. 南京: 东南大学,2006.
[10]Aibe Valter Yoshihiko. Water mass measurement m parts for volume standard calibration in tlie key comparison[C]//FLOMEKO.14th International Flow Measurement Conference 2007. Johannesburg, South Africa, 2007: 145-148.
[11]Batista E, Pinto L, Filipe E, et al. Calibration of micropipettes:Test methods and uncertainty analysis[J]. Measurement, 2007, 40(3):338-342.
[12]刘子勇. 容量计量[M].北京:中国计量出版社,2009.
[13]Batista E, Arias R, WANG J T. Final report on BIPM/CIPM key comparison CCM.FF-K4.2.2011:Volume comparison at 100 μL—Calibration of micropipettes (piston pipettes)[J]. Metrologia, 2013, 50(1A):07003.
[14]陈超云, 王金涛, 李之昊, 等. 基于静力称重法的玻璃浮计校准方法研究[J]. 计量学报, 2017, 38(2):197-199.