Research of Constant Temperature and Velocity Water Bath for Calibration of Temperature Sensors Dynamic Response
ZHENG Wei1,LIU Bo1,WANG Zhiyuan2
1. Shanghai Institute of Measurement and Testing Technology, Shanghai 201203, China
2. School of Energy and Power Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
Abstract:Dynamic response is a key performance of temperature sensors in important and safety occasions. The constant temperature and velocity water bath is the core equipment for the calibration of temperature sensors dynamic response. Due to the slow heating rate in typical annular rotating water bath, a novel electrically heated annular rotating water bath was developed based on the precision sliding ring, and the temperature fluctuations and velocity fields of the novel bath were tested. The results show that the heating up time is shortened, the temperature fluctuation is significantly less than ±1.0% of the step temperature, the vertical velocity uniformity is significantly better than 10% of the novel bath. The time constants of a sheathed resistance thermometer and a sheathed thermocouple were measured by the existing jet rotating water bath and the novel bath. The deviations from the existing bath results are 4.1% and 7.5%, respectively, these meet the engineering requirement. The novel bath provides good constant temperature and velocity environmental conditions for the dynamic response calibration of temperature sensors.
郑伟,刘波,王志远. 温度传感器动态响应校准恒温恒速水槽研究[J]. 计量学报, 2024, 45(9): 1257-1261.
ZHENG Wei,LIU Bo,WANG Zhiyuan. Research of Constant Temperature and Velocity Water Bath for Calibration of Temperature Sensors Dynamic Response. Acta Metrologica Sinica, 2024, 45(9): 1257-1261.
Standard test methods for testing industrial resistance thermometers: ASTM E644-11(2019) [S]. 2019.
[14]
核电厂安全级电阻温度探测器的质量鉴定: GB/T 25838-2010 [S]. 2010.
[1]
LIU Y, MITSUTAKE Y, MONDE M. Development of fast response heat transfer measurement technique with thin-film thermocouples [J]. International Journal of Heat and Mass Transfer, 2020, 162: 120331.
SUN H J, LI W J, LI J Q, et al. A thermocouple time constant test system [J]. Journal of China University of Metrology, 2017, 28(2): 146-152.
LIU B, WANG C Y, ZHU P F. Uncertainty evaluation on temperature sensors dynamic response based on water facility [J]. Acta Metrologic Sinica, 2020, 41(10): 1240-1244.
WANG P P, ZHAO S W, MENG W R, et al. Research on measurement uncertainty of dynamic response characteristics of ocean temperature sensor based on LSM [J]. Acta Metrologic Sinica, 2023, 44(6): 886-892.
[13]
HASHEMIAN H M, JIANG J. Accuracy in temperature sensor response time estimation for new nuclear reactor designs [J]. International Journal of Nuclear Energy Science and Technology, 2011, 6(1): 17-29.
[2]
TALER D, SOBOTA T, JAREMKIEWICZ M, et al. Influence of the thermometer inertia on the quality of temperature control in a hot liquid tank heated with electric energy [J]. Energies, 2020, 13(15): 4039.
WEN J H, WANG J, YANG L, et al. Response time of microfiber temperature sensor in liquid environment [J]. IEEE Sensors Journal, 2020, 20 (12): 6400-6407.
RUPNIK K, KUTIN J, BAJSIC′ I. Identification and prediction of the dynamic properties of resistance temperature sensors [J]. Sensors and Actuators A: Physical, 2013, 197: 69-75.
ZHAO J. Analysis of key influencing factors of time constant of gas flow temperature sensors [J]. Acta Metrologic Sinica, 2022, 43(12): 1581-1586.
ZHANG L, PENG Y, WANG X. Simulation and optimization design of aviation nickel resistance temperature sensor [J]. Journal of Aerospace Power, 2021, 36(2): 440-448.
DINIZ A C G C, DEALMEIDA M E K, VIANNA J N S, et al. Methodology for estimating measurement uncertainty in the dynamic calibration of industrial temperature sensors [J]. Journal of the Brazilian Society of Mechanical Sciences and Engineering, 2017, 39(3): 1053-1060.
JIA C, CAI J, XIONG Z H. Study on dynamic characteristics of temperature sensors under different heat transfer modes [J]. Acta Metrologica Sinica, 2020, 41(5): 563-566.