Abstract:50Hz~10kHz voltage ratio calibration system based on digital acquisition device is proposed. The system is composed of voltage conversion unit and digital acquisition device. The channel switching technology is used to greatly improve the accuracy of voltage ratio with the digital acquisition device. The voltage conversion unit is composed of cascade inductive voltage divider, isolation transformer and high & low voltage precision voltage follower. In this part, the turn-back winding process and so on are proposed to reduce the additional error caused by the stray parameters and the iron core material, and a two-stage isolation transformer with twisted pair structure winding is designed. The amplitude error of the two-stage transformer within the 10kHz range shall not exceed 5×10-6, and the phase error shall not exceed 10μrad. Precise low-voltage and high-voltage composite voltage followers are designed to compensate the phase response by using the characteristic matching of the double-package operational amplifier, which significantly improves the phase characteristics of the voltage follower, within the 10kHz range, the amplitude error of the follower shall not exceed 10×10-6, and the phase error shall not exceed 20μrad.The measurement results indicate that the calibration system has high measurement accuracy, and the measurement deviation at 50Hz is not greater than 1×10-6 and 1μrad. Within the 10kHz range, the ratio error deviation is not more than 80×10-6, and the phase error deviation is not more than 60μrad, which can meet the voltage ratio error measurement requirements in different frequency scenarios.
GHASSEMI F. Harmonic Voltage Measurements Using CVTs[J]. IEEE Transactions on Power Delivery, 2005, 20(1): 443-449.
[7]
CROTTI G, GALLO D, GIORDANO D, et al. Frequency Response of MV Voltage Transformer Under Actual Waveforms[J]. IEEE Transactions on Instrumentation and Measurement, 2017, 66(6): 1146-1154.
[9]
CROTTI G, D′AVANZO G, LANDI C, et al. Evaluation of Voltage Transformers Accuracy in Harmonic and Interharmonic Measurement[J]. IEEE Open Journal of Instrumentation and Measurement, 2022, 1: 1-10,
GAO H L, LI Q L. Harmonic Transfer Characteristic of Capacitor Voltage Transformer[J]. Power System Technology, 2013, 37(11): 3125-3130.
JIANG C Y, LIU J, WANG X, et al. Research on test method of harmonic measurement error ofinductive voltage transformer [J]. China Measurement & Test, 2021, 47(3): 163-168.
[15]
MOHNS E, MEISNER J, ROEISSLE G, et al. A Wideband Current Transformer Bridge[J]. IEEE Transactions On Instrumentation And Measurement, 2014, 63(10): 2322-23229.
ZHOU F, LEI M, WANG L R, et al. A Transformer Voltage Series Method Based on Three-Port Network[J]. Acta Metrologica Sinica, 2021, 42(4): 401-408.
JIN H B, LI X Z, WANG Q J, et al. Induction Ratio Bridge Balance Method-based Synchro/resolver Standard Calibration Device[J]. Acta Metrologica Sinica, 2021, 42(9): 1200-1205.
[22]
LOQUINARIO J. Composite Amplifiers: High Output Drive Capability with Precision[EB/OL]. https://www.analog.com/media/en/analog-dialogue/volume-53/number-3/composite-amplifiers—high-output-drive-capability-with-precision.pdf.
LIU G, SUN L P, WANG X G, et al. Improvement of Core Loss Calculation Method and Simulation Application under Sinusoidal and Harmonic Excitations[J]. Transactions of China Electrotechnical Society, 2018, 33(21): 4189-4198.
JIANG C Y, LIU H, ZHOU F, et al. The development of an active capacitive voltage divider[J]. Electrical Measurement & Instrumentation, 2019, 56(6): 148-152.
JIANG C Y, ZHOU F, YANG S H, et al. Development of Broadband Capacitive Voltage Divider[J]. High Voltage Apparatus, 2017, 53(1): 151-156.
[8]
CROTTI G, D′AVANZO G, LETIZIA P S, et al. Measuring Harmonics With Inductive Voltage Transformers in Presence of Subharmonics[J]. IEEE Transactions on Instrumentation and Measurement, 2021, 70: 1-13.
LI H Q, JIA Y F, CHEN X, et al. Experimental Research on Excitation Characteristics and Loss of Steel Sheet under Interharmonic Excitation[J]. Transactions of China Electrotechnical Society, 2018, 33(18): 4192-4198.
[11]
DAVIS P, WRIGHT P, KING J, et al. Voltage Transformer Harmonic Characteristics For Distribution Power Quality Monitoring[C]// CIRED 2021—The 26th International Conference and Exhibition on Electricity Distribution, Online Conference.2021: 820-824.
PENG S X, SONG Y H, MA Y X. R&D in double level potential transformers of 35kV with an accuracy of 0. 001 and 110/3kV with an accuracy of 0. 002[J]. Electrical Equipment, 2005(7): 10-16.
XIAO X N, LIAO K Y, TANG S H, et al. Development of Power-Electronized Distribution Grids and the New Supraharmonics Issues[J]. Transactions of China Electrotechnical Society, 2018, 33 (4): 708-720.
[10]
TOSCANI S, FAIFER M, FERRERO A, et al. Compensating Nonlinearities in Voltage Transformers for Enhanced Harmonic Measurements: The Simplified Volterra Approach[J]. IEEE Transactions on Power Delivery, 2021, 36(1): 362-370.
[14]
岳长喜, MOHNS E, 周峰, 等. 一种电压比例精密测量方法[J], 电测与仪表, 2013, 50(3): 37-40.
[19]
AWAN S, KIBBLE B, SCHURR J. Coaxial Electrical Circuits for Interference-Free Measurements (IET Electrical Measurement)[J]. IET Digital Library,2011,DOI:10.1049/PBEL013E.