1.Fujian Institute of Metrology,Fuzhou,Fujian 350003,China
2.Anhui Institute of Metrology,Hefei, Anhui 230051,China
3.National Institute of Metrology,Beijing 100029,China
Abstract:The half-value layer measurement is required in various diagnostic X-ray radiation source verification regulations or calibration specifications, and half value layer measuring instruments are also widely used, but the calibration method of half value layer measuring instruments has not been clear. The main influencing factors in the calibration of the half value layer measuring instrument are discussed. The results show that the detector angle has the greatest influence on the measurement results. Therefore, the half value layer measuring instrument to be calibrated should be placed in the center of the test plane during calibration, and the angle deviation should not exceed ±1°.
Sui S S, Wei J, Shi Q. The Calcylation and Application of Mass Attenuation Conefficients[J]. Spectroscopy and Spectral Analysis, 1996, 16(2): 85-89.
Wang T Q, Wei X D, Li H J, et al. Calculation of X-ray Attenuation and Energy Deposition [J]. Atomic Energy Science and Technology, 2007, 41 (4): 399-403.
Li Y P, Wu L P, Xie W, et al. The Study on the Measurement of X-Rays Mass Absorption Coefficients[J]. Journal of Sichuan University (Natural Science Edition), 2005, 42(2): 343-346.
Jin Y B, Jiang N. Further Research on HVL Measurement Methods[J]. China Medical Device Information, 2014(3): 51-53.
[9]
JJG1078—2012 医用数字摄影(CR、DR)系统X射线辐射源[S]. 2012.
Zou J X, Wang P D, Li X S, et al. Measurement of half-value layer for medical diagnostic X-ray[J]. Shanghai Measurement and Testing, 2020, 47(2):17-20.
Hao Y M, Lü Y Z, Li D H, et al. Initial Exploration of Half-value Layer Determination Using Thermoluminescent Dosimeter[J]. Acta Metrologica Sinica, 2021, 42(6): 811-814.
Wang P, Xing L T, Jiang W. Measurement and Research for the Filtered X-ray Radiation Quality of Low Air Kerma Rate Series[J]. Acta Metrologica Sinica, 2021, 42(4): 521-525.
Li D H, Guo B, Huang J W, et al. Study on simulation and measurement of diagnostic X-ray radiation qualities [J]. Chin J Radiol Med Prot, 2017, 37(8): 614-617.
Li B, Wu J J, Fang F, et al. The Researchand Measurement of Medical Diagnostic X-ray Beam Half-Value Layer and Homogeneous Coefficient [J]. Nuclear Electronics & Detection Technology, 2012, 32 (8): 946-952.