2025年05月28日 星期三 首页   |    期刊介绍   |    编 委 会   |    投稿指南   |    期刊订阅   |    统合信息   |    联系我们
计量学报  2022, Vol. 43 Issue (7): 869-876    DOI: 10.3969/j.issn.1000-1158.2022.07.06
  热学计量 本期目录 | 过刊浏览 | 高级检索 |
PPRTs偏差方程外推至-189.3442~156.5985℃温区的研究
王博阳1,曾凡超2,黄安贻1,孙建平3,傅承玉2
1.武汉理工大学,湖北 武汉 430070
2.湖北省计量测试技术研究院,湖北 武汉 430223
3.中国计量科学研究院,北京 100029
Research on Extrapolation of PPRTs Deviation Equation to -189.3442~156.5985℃
WANG Bo-yang1,ZENG Fan-chao2,HUANG An-yi1,SUN Jian-ping3,FU Cheng-yu2
1. Wuhan University of Technology, Wuhan, Hubei 430070, China
2. Hubei Instiute of Measurement and Testing, Wuhan, Hubei 430223, China
3. National Institute of Metrology, Beijing 100029, China
全文: PDF (4304 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 由于ITS-90定义固定点数量少且温度间隔大,基于偏差方程的外推是解决超出温区范围的温度计标定问题的有效方法。实验对16支精密铂电阻温度计在超出温区范围标定,验证了固定点法和比较法偏差方程从窄温区范围外推至-189.3442~156.5985℃温区的可行性。实验结果表明:-38.8344~0.01℃温区新偏差方程外推至 -189.3442℃ 的平均最大差值为5.2mK;0~29.7646℃温区偏差方程外推至156.5985℃的平均最大差值为 2.0mK;基于比较法的偏差方程在-189.3442~156.5985℃温区的平均差值小于3.3mK。3种方程均提高了精密铂电阻温度计在超出温区范围的外推精度,为星载黑体辐射源的量值溯源提供了数据支撑。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
王博阳
曾凡超
黄安贻
孙建平
傅承玉
关键词 计量学90国际温标温度固定点偏差方程外推精密铂电阻温度计量值溯源    
Abstract:The traceability of precision platinum resistance thermometers (PPRTs) can be improved by using the fixed-points method defined by the International Temperature Scale of 1990 (ITS-90). Due to the small number of fixed-points and the large temperature interval in ITS-90, the extrapolation based on deviation equation is an effective method to solve the problem of thermometer calibration beyond the temperature range. In the experiment, 16 PPRTs were calibrated in the temperature range of -189.3442~156.5985℃ to verify the extrapolation feasibility of the fixed-points method deviation equation and comparison method deviation equation.The experimental results showed that the average maximum difference using the new deviation equation from -38.8344~0.01℃ was 5.2mK when extrapolating to -189.3442℃; the average maximum difference using the deviation equation from 0~29.7646℃ was 2.0mK when extrapolating to 156.5985℃; and the average difference using comparison method deviation equation was less than 3.3mK.All the three equations improved the extrapolation accuracy of PPRTs out of the fixed-points temperature range and provided supporting data for the traceability of spaceborne blackbody radiation source.
Key wordsmetrology    ITS-90    temperature fixed point    extrapolation of deviation equation    PPRT    metrological traceability
收稿日期: 2021-04-07      发布日期: 2022-07-18
PACS:  TB942  
基金资助:国家重点研发计划(2018YFB0504700;2018YFB0504702)
通讯作者: 曾凡超(1991-),湖北武汉人,湖北省计量测试技术研究院工程师,主要从事热工计量研究。Email:zengfanchao3@163.com     E-mail: zengfanchao3@163.com
作者简介: 王博阳(1997-),四川绵阳人,武汉理工大学研究生,主要研究方向为计量技术及精密仪器。Email:WHUTwangboyang@163.com
引用本文:   
王博阳,曾凡超,黄安贻,孙建平,傅承玉. PPRTs偏差方程外推至-189.3442~156.5985℃温区的研究[J]. 计量学报, 2022, 43(7): 869-876.
WANG Bo-yang,ZENG Fan-chao,HUANG An-yi,SUN Jian-ping,FU Cheng-yu. Research on Extrapolation of PPRTs Deviation Equation to -189.3442~156.5985℃. Acta Metrologica Sinica, 2022, 43(7): 869-876.
链接本文:  
http://jlxb.china-csm.org:81/Jwk_jlxb/CN/10.3969/j.issn.1000-1158.2022.07.06     或     http://jlxb.china-csm.org:81/Jwk_jlxb/CN/Y2022/V43/I7/869
京ICP备:14006989号-1
版权所有 © 《计量学报》编辑部
地址:北三环东路18号(北京1413信箱)  邮编:100029 电话:(010)64271480
本系统由北京玛格泰克科技发展有限公司设计开发  技术支持:support@magtech.com.cn