真空汞固定点黑体辐射源的设计与研制

李凯, 郝小鹏, 宋健, 孙建平, 胡朝云, 刘洋, 扈又华, 杨延龙

计量学报 ›› 2020, Vol. 41 ›› Issue (4) : 413-418.

PDF(1332 KB)
PDF(1332 KB)
计量学报 ›› 2020, Vol. 41 ›› Issue (4) : 413-418. DOI: 10.3969/j.issn.1000-1158.2020.04.004
热学计量

真空汞固定点黑体辐射源的设计与研制

  • 李凯1,2, 郝小鹏2, 宋健2, 孙建平2, 胡朝云2, 刘洋2,3, 扈又华1,2, 杨延龙1,2
作者信息 +

Design and Development of Vacuum Mercury Fixed PointBlackbody Radiation Source

  • LI Kai1,2, HAO Xiao-peng2, SONG Jian2, SUN Jian-ping2, HU Chao-yun2, LIU Yang2,3, HU You-hua1,2, YANG Yan-long1,2
Author information +
文章历史 +

摘要

介绍了中国计量科学研究院研制的真空汞固定点黑体辐射源的结构、工作原理、性能测试结果和不确定度分析。真空汞固定点黑体辐射源灌注的是纯度为99.9999%的高纯汞,黑体空腔开口直径为25mm,空腔内径为28mm,深度为260mm,表面喷涂了NEXTEL 811-21高发射率涂层,采用基于蒙特卡罗黑体发射率仿真计算的方法,计算了黑体空腔在波长为8~14μm的发射率,结果优于0.9999;在真空环境下,测试了真空汞固定点黑体辐射源的温坪曲线和重复性等主要技术指标,结果表明真空汞固定点黑体辐射源温坪稳定性优于2mK,多次重复性优于1mK;分析了真空汞固定点黑体辐射源的不确定度来源,其合成标准不确定度为16mK。

Abstract

The structure, working principle, performance test results and uncertainty analysis of the vacuum mercury fixed point blackbody radiation source developed by NIM, China, were introduced. The perfusion metal of the vacuum mercury fixed point blackbody was high purity mercury with a purity of 99.9999%. The opening diameter of the blackbody cavity is 25 mm, the inner diameter of the cavity is 28 mm, the depth is 260mm, and the surface is coated with high emissivity coating from NEXTEL 811-21. The Monte Carlo calculation method was used to calculate the blackbody emissivity at the wavelength of 8~14μm. The emissivity is higher than 0.9999. In the vacuum environment, the main technical indicators such as the temperature curve and repeatability of the temperature of the mercury fixed point blackbody were tested. The results show that the temperature stability of the vacuum mercury fixed point blackbody is better than 2mK, and the repeatability is better than 1mK. The uncertainty source of the vacuum mercury fixed point blackbody is analyzed, the synthetic standard uncertainty is better than 16mK.

关键词

计量学 / 汞固定点 / 黑体辐射源 / 相变温坪复现 / 不确定度 / metrology / mercury fixed point / blackbody radiation source / reproducibility of phase transition plateau / uncertainty / 计量学 / 汞固定点 / 黑体辐射源 / 相变温坪复现 / 不确定度

Key words

metrology / mercury fixed point / blackbody radiation source / reproducibility of phase transition plateau / uncertainty

引用本文

导出引用
李凯, 郝小鹏, 宋健, 孙建平, 胡朝云, 刘洋, 扈又华, 杨延龙. 真空汞固定点黑体辐射源的设计与研制[J]. 计量学报. 2020, 41(4): 413-418 https://doi.org/10.3969/j.issn.1000-1158.2020.04.004
LI Kai, HAO Xiao-peng, SONG Jian, SUN Jian-ping, HU Chao-yun, LIU Yang, HU You-hua, YANG Yan-long. Design and Development of Vacuum Mercury Fixed PointBlackbody Radiation Source[J]. Acta Metrologica Sinica. 2020, 41(4): 413-418 https://doi.org/10.3969/j.issn.1000-1158.2020.04.004
中图分类号: TB942   

参考文献

1 郝小鹏, 宋健, 孙建平, 等. 风云卫星红外遥感亮度温度国家计量标准研究[J].光学精密工程, 2015, 23(7): 1845-1851. HaoX P, SongJ, SunJ P, et al. Vacuum radiance temperature national standard facility for infrared remote sensors of Chinese Fengyun meteorological satellites [J]. Optics and Precision Engineering, 2015, 23(7): 1845-1851.
2 AndersonJ G, DykemaJ A, GoodyR M, et al. Absolute, spectrally-resolved, thermal radiance: a benchmark for climate monitoring from space [J]. Journal of Quantitative Spectroscopy & Radiative Transfer, 2004, 85(3): 367-383.
3 DykemaJ A, AndersonJ G. A methodology for obtaining on-orbit SI-traceable spectral radiance measurements in the thermal infrared [J]. Metrologia, 2006, 43(6), 287-293.
4 DechS, KleinD. Challenges of Earth Observation for Global Change Monitoring [C]//Earsel Symposium. 2016.
5 宋健, 郝小鹏, 丁雷,等. 真空低背景红外高光谱亮温计量标准装置研制[J].红外与激光工程, 2019, 48(10): 1004001. SongJ, HaoX P, DingL, et al. Set up Vacuum Infrared Hyperspectral Radiance Temperature Standard Facility [J]. Infrared and Laser Engineering, 2019, 48(10): 1004001.
6 MonteC, GutschwagerB, MorozovaS P, et al. Radiation Thermometry and Emissivity Measurements Under Vacuum at the PTB [J]. International Journal of Thermophysics, 2009, 30(1): 203-219.
7 CarterA C, DatlaR U, JungT M, et al. Low-background temperature calibration of infrared blackbodies [J]. Metrologia International Journal of Scientific Metrologia, 2006, 43(2): S46-S50.
8 FowlerJ B, JohnsonB C, RiceJ P, et al. The new cryogenic vacuum chamber and black-body source for infrared calibrations at the NIST's FARCAL facility [J]. Metrologia, 1998, 35(4): 323-327.
9 MekhontsevS, KhromchenkoV, ProkhorovA, et al. Establishing a New NIST Facility for the Primary Realization of both Spectral Radiance and Reflectance in the Mid-and Far-Infrared[C]//NEWRAD, USA, Maui Hawaii, 2011, 26(7): 32-33.
10 KhromchenkoV B, MekhontsevS N, HanssenL M. Design and Evaluation of Large-Aperture Gallium Fixed-Point Blackbody[J]. International Journal of Thermophysics, 2009, 30(1): 9-19.
11 IvanovV S, LisianskyB E, MorozovaS P, et al. Medium-background radiometric facility for calibration of sources or sensors[J]. Metrologia, 2000, 37(5): 599-602.
12 许敏, 郝小鹏, 宋健, 等. 190~340K真空标准黑体辐射源研制[J]. 计量学报, 2015, 36(zl): 50-54. XuM, HaoX P, SongJ, et al. Research of Vacuum Radiance Temperature Standard Blackbody Source from 190K to 340 K[J]. Acta Metrologica Sinica, 2015, 36(zl): 50-54.
13 舒心, 郝小鹏, 宋健, 等. 100~400K真空红外亮温标准黑体辐射源研制[J]. 计量学报, 2019, 40(1): 13-19. ShuX, HaoX P, SongJ, et al. Research of 100 ~ 400 K Vacuum Infrared Radiance Temperature Standard Blackbody Source [J]. Acta Metrologica Sinica, 2019, 40(1): 13-19.
14 胡朝云,郝小鹏,宋健,等. 红外高光谱大气探测仪星载固定点黑体辐射源的研制[J].计量学报, 2019, 40(2): 232-239. HuC Y,HaoX P,SongJ, WENPing,et al. Development of Blackbody Radiation Sources at Fixed Point on Satellite of Infrared Hyperspectral Atmospheric Detector[J].Acta Metrologica Sinica, 2019, 40(2): 232-239.
15 AdibekyanA, KononogovaE, MonteC, et al. High Accuracy Emissivity Data on the Coatings Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black[J]. International Journal of Thermophysics, 2017, 38(6): 89.
16 宋健, 郝小鹏, 原遵东, 等. 基于控制环境辐射的黑体辐射源发射率测量方法研究[J]. 中国激光, 2015, 42(9): 269-275. SongJ, HaoX P, YuanZ D, et al. Blackbody Source Emissivity Measurement Method Based on Controlling Surroundings Radiation[J]. Chinese Journal of Lasers. 2015, 42(9): 269-275.
17 宋健, 郝小鹏, 原遵东, 等 . 基于控制环境辐射发射率测量方法的超黑涂层发射率特性研究[J].计量学报, 2015, 36(zl): 24-27. SongJ, HaoX P, YuanZ D, et al. Research on Ultra Black Paint Emissivity of the Emissivity Measure Method Based on Controlling Surroundings Radiation [J]. Acta Metrologica Sinica, 2015, 36(zl): 24-27.
18 SaundersP, FischerJ, SadliM, et al. Uncertainty Budgets for Calibration of Radiation Thermometers below the Silver Point[J]. International Journal of Thermophysics, 2008, 29(3): 1066-1083.
19 原遵东, 陈桂生 . 发射率影响修正近似公式及其适用性分析[J].计量技术, 2013, (4): 7-10.

基金

国家重点研发计划(2018YFB0504700,2018YFB0504702);国家自然科学基金(11475162)