1. Changchun University of Science & Technology, Changchun, Jilin 130022, China
2. National Institute of Metrology, Beijing 100029, China
3. Xi‘an Polytechnic University, Xi'an, Shaanxi 710048, China
Abstract:The IR-RST standard radiation thermometers manufactured by CHINO Company, Limited, Japan, which can not display the temperature in real-time because the display unit hadn’t been designed. Therefore, an electrical measurement display device was designed which can measure the voltage, display the temperature and supply 24V DC voltage for this kind of infrared radiation thermometer. Calibrated a against high-level voltage power source and a voltmeter, the correction coefficients of the display instrument were got, the stabilitywas evaluated as well, the maximum relative error of voltage measurement can be achieved with in 0.03%. Assembled with a 0.65μm IR-RST radiation thermometer, it was calibrated against with a TG HT-9500 type high-temperature furnace as the source, the accuracy was better than 0.03% confirmed by the experimental results, and the extended uncertainty of the calibration system was 0.62℃ at 1 300℃.
盛山菊,卢小丰,唐小洁,宦克为,王雪莲. 红外辐射温度计专用数字表的研制及评价[J]. 计量学报, 2019, 40(1): 20-24.
SHENG Shan-ju,LU Xiao-feng,TANG Xiao-jie,HUAN Ke-wei,WANG Xue-lian. Development and Evaluation of Digital Meter for an Infrared Radiation Thermometer. Acta Metrologica Sinica, 2019, 40(1): 20-24.
1]Yuan G. Optical system and focusing structure for infrared thermometer: US 8870452[P]. 2014-10-28.
[2]原遵东, 邢波, 柏成玉, 等. 有效亮度温度测量中发射率影响的修正[J]. 计量学报, 2014, 35(6): 578-582.
Yuan Z D, Xing B, Bai C Y, et al. Correcting effective radiance temperature measurement for emissivity effect[J]. Acta Metrologica Sinica, 2014, 35(6): 578-582.
[3]张文琦, 房落凤, 田鹏飞, 等. 红外辐射测温系统设计及 SSE 的影响研究[J]. 仪表技术与传感器, 2015,(2): 71-73.
Zhang W Q, Fang L F, Tian P F, et al. Design of infrared thermometry system and research on effecta of SSE[J].Instrument Technique and Sensor, 2015, (2): 71-73.
[4]黄燊彦, 郝小鹏, 王景辉, 等. 铯, 钠热管标准黑体辐射源研制及性能评价[J]. 计量学报, 2014, 35(2): 120-124.
Huang S Y, HAO X P,Wang J H, et al. Development and characteration of cesium, sodium heat pipe standard blackbody radiation sources [J]. Acta Metrologica Sinica, 2014, 35(2): 120-124.
[5]孙安青. PIC单片机实用C语言程序设计与典型实例[M]. 北京:中国电力出版社, 2008.
[6]原遵东, 郝小鹏, 王景辉,等. 黑体辐射源的多波长有效亮度温度校准和不同溯源方式特点分析[J]. 计量学报, 2017, 38(2):135-140.
Yuan Z D, Hao X P, Wang J H,et al. Multi-wavelength effective radiance temperature calibration and characteristics analysis of traceable approaches of the blackbody radiation sources[J]. Acta Metrologica Sinica, 2017, 38(2): 135-140.
[7]栾海峰, 卢小丰, 王景辉, 等. 以高温黑体为比较源的 0.9 μm 标准高温计校准方法研究[J]. 计量学报, 2015, 36(z1): 20-23.
Luan H F, Lu X F,Wang J H, et al. Research on the calibration of 0.9 μm radiation pyrometer based on high temperature blackbody[J]. Acta Metrologica Sinica, 2015, 36(z1):20-23.
[8]卢小丰,原遵东,董伟,等.应用高温固定点校准精密光电高温计[J].计量学报,2017,38(5):584-588.
Lu X F, Yuan Z D, Dong W, et al. Calibrating pyrometers with high temperature fixed points[J]. Acta Metrologica Sinica, 2017, 38(5): 584-588.
[9]Lee R D. The NBS Photoelectric Pyrometer and its Use in Realizing the International Practical Temperature Scale above 1 063 ℃[J]. Metrologia, 1966, 2(4):150.
[10]Sakuma F, Hattori S. Establishing a Practical Temperature Standard by Using a Narrow-band Radiation Thermometer with a Silicon Detector[C]//Temperature:Its Measurement and Control in Science and Industry. New York,1982.
[11]倪育才.实用测量不确定度评定[M].北京:中国计量出版社,2004:79-84.
[12]卢小丰, 原遵东, 王景辉,等. 以“固定点-高温计”方式传递961.78℃以上国际温标[J]. 计量学报, 2014, 35(3):193-197.
Lu X F, Yuan Z D, Wang J H, et al. Disseminating the international temperature scale above 961.78 ℃ in the “fixed point-pyrometer” method[J]. Acta Metrologica Sinica, 2014, 35(3): 193-197.