Method for the tractability of Quantity Value of Sonic Nozzles with Stagnation Pressure as Low as 10 kPa
LI Chun-hui1,SU Jin-you2,YUAN Shi-hui2,GAO Shan1,3,ZHOU Li-yuan1
1. National Institute of Metrology, Beijing 100029, China
2. Key Laboratory of High-altitude Simulation Technology, Mianyang, Sichuan 621024, China
3. College of Quality and Technical Supervision, Hebei University, Baoding, Hebei 071002, China
Abstract:In order to solve the problem that the existing gas flow measurement calibration technology can not achieve the direct calibration of the air engines gas flow under low pressure operating conditions and the traceability of the quantity value, the method of direct calibration and quantity tracing of sonic nozzles under low stagnation pressure was studied. Firstly, four sonic nozzles with nominal throat diameter of 40 mm were tested under atmospheric pressure using the gas flow reference device of pVTt method. Secondly, based on five standard sonic nozzles, the real flow calibration of sonic nozzles with a minimum of 10 kPa stagnation pressure is achieved by using a series of sonic nozzles. Finally, the discharge coefficient measured is linearly matched with the square root reciprocal of the Reynolds number in the throat of nozzle. Based on the linear fitting results, the feasibility and accuracy of the method are verified.
李春辉,苏金友,袁世辉,高山,周立媛. 滞止压力低至10kPa的音速喷嘴量值溯源方法探析[J]. 计量学报, 2018, 39(6A): 104-.
LI Chun-hui,SU Jin-you,YUAN Shi-hui,GAO Shan,ZHOU Li-yuan. Method for the tractability of Quantity Value of Sonic Nozzles with Stagnation Pressure as Low as 10 kPa. Acta Metrologica Sinica, 2018, 39(6A): 104-.
[1]Smith R E, Matz R J. A theoretical method of determining discharge coefficients for Venturi operating at critical flow conditions[J]. Journal of Basic Engineering, 1962, 84(4): 434-445.
[2]Stratford B S. The Calculation of the Discharge Coefficient of Profiled Choked Nozzles and the Optimum Profile for Absolute Air Flow Measurement[J]. Journal of the Royal Aeronautical society, 1964: 237-244.
[3]Vallet J P, Windenberger C, Vulovic F, et al. Improvement of thermodynamic calculations used for the flow rate of sonic nozzles[C]//4th International Symposium on Fluid Flow Measurement, Colorado USA, 2000.
[4]Mickan B, Kramer R H, Dopheide D, et al. Comparisons by PTB, NIST and LNE-LADG in air and natural gas with critical Venturi nozzles agree within 0. 05%[C]//IMEKO World Congress 18th, Queretaro, MX, 2006.
[5] U. S. National standards for High Pressure Natural Gas Flow Measurement[S]. 2008.
[6]Ishibashi M. Proposal of Fluid Dynamical Standard (FDS) for Gas Flow-rate[C]// ASME FEDSM, 2002: 1-13.
[7]Tang S S. Discharge coefficient correlations for critical flow nozzles and their dependence on Reynolds numbers[Z]. Princeton University, 1969.
[8]Geropp D. Laminare Grenzschichten in Ebenen und Rotation-Symmetrischen Lavaldüesen[J]. Deutsche Luft-Und Raumfart Forschungsbericht, 1971: 71-90.
[9]Arnberg B T, Britton C L, Seidl W F. Discharge coefficient correlations for circular-arc Venturi flowmeters at critical (sonic) flow[J]. Journal Fluids Engineering, 1974, 96(2):111-123.
[10]ISO 9300, Measurement of Gas Flow by Means of Critical Flow Venturi Nozzles [S]. 2005.