计量学报  2017, Vol. 38 Issue (6): 681-685    DOI: 10.3969/j.issn.1000-1158.2017.06.05
  光学计量 本期目录 | 过刊浏览 | 高级检索 |
400~1700 nm绝对光谱响应度校准和量值传递
史学舜1,徐楠2,刘长明1,甘海勇2,李健军3,林延东2,陈坤峰1
1. 中国电子科技集团公司 第四十一研究所, 山东 青岛 266555
2. 中国计量科学研究院, 北京 100029
3. 中国科学院 通用光学定标与表征技术重点实验室, 安徽 合肥 230031
Calibration and value transfer of absolute spectral responsivity between 400nm and 1700nm
SHI Xue-shun1,XU Nan2,LIU Chang-ming1,GAN Hai-yong2,LI Jian-jun3,LIN Yan-dong2,CHEN Kun-feng1
1. The 41st Research Institute of CETC, Qingdao, Shandong 266555, China
2. National Institute of Metrology, Beijing 100029, China
3. Key Laboratory of Optical Calibration and Characterization of CAS, Hefei, Anhui 230031, China
全文: PDF (436 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 介绍了高准确度光辐射功率校准原理和方法,利用低温辐射计作为主标准器,以陷阱探测器作为传递标准,激光器作为光源,通过激光功率稳定装置,校准了硅陷阱探测器和铟镓砷陷阱探测器的绝对光谱响应度。选取476.1, 488, 514.7, 521, 568, 632.8, 647.1, 785, 852, 980, 1064, 1550,nm共12条谱线完成了校准实验,绝对光谱响应度测量不确定度均优于0.05%。通过量子效率模型得出了硅陷阱探测器的绝对光谱响应曲线。利用InGaAs陷阱探测器分立波长点的绝对光谱响应度与相对光谱响应曲线进行了验证分析。结果表明,2种陷阱探测器均可用作传递标准进行高准确度的可见光和近红外光辐射功率校准和量值传递。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
史学舜
徐楠
刘长明
甘海勇
李健军
林延东
陈坤峰
关键词 计量学绝对光谱响应度绝对低温辐射计陷阱探测器不确定度    
Abstract:The principle and methods of optical radiation measurement was illustrated. The absolute spectral responsivity of Si trap detectors and InGaAs trap detectors were calibrated with the absolute cryogenic radiometer at eleven wavelengths between 400 nm and 1700 nm, which are 476.1, 488, 514.7, 521, 568, 632.8, 647.1, 785, 852, 980, 1064 and 1550 nm, respectively, with measurement uncertainties better than 0.05%. The absolute spectral responsivity curves of Si trap detectors were calculated by quantum efficiency model. The absolute responsivity of three wavelengths was applied to get absolute spectral responsivity from the relative spectral responsivity of InGaAs detector. Results show that the two types of trap detectors can be used to calibrate and transfer optical radiation value in the visible and near infrared spectral range.
Key wordsmetrology    absolute spectral responsivity    absolute cryogenic radiometer    trap detector    uncertainty
收稿日期: 2015-12-25      发布日期: 2017-09-27
PACS:  TB96  
基金资助:国家“863”计划(2015AA123702); 国家技术基础项目(J312013A001)
通讯作者: 史学舜     E-mail: xshshi@163.com
作者简介: 史学舜(1975-),男,山东临朐人,中国电子科技集团公司第四十一研究所高级工程师, 主要从事光辐射参数与光电探测器参数计量技术研究。xshshi@163.com
引用本文:   
史学舜,徐楠,刘长明,甘海勇,李健军,林延东,陈坤峰. 400~1700 nm绝对光谱响应度校准和量值传递[J]. 计量学报, 2017, 38(6): 681-685.
SHI Xue-shun,XU Nan,LIU Chang-ming,GAN Hai-yong,LI Jian-jun3,LIN Yan-dong,CHEN Kun-feng. Calibration and value transfer of absolute spectral responsivity between 400nm and 1700nm. Acta Metrologica Sinica, 2017, 38(6): 681-685.
链接本文:  
http://jlxb.china-csm.org:81/Jwk_jlxb/CN/10.3969/j.issn.1000-1158.2017.06.05     或     http://jlxb.china-csm.org:81/Jwk_jlxb/CN/Y2017/V38/I6/681
京ICP备:14006989号-1
版权所有 © 《计量学报》编辑部
地址:北三环东路18号(北京1413信箱)  邮编:100029 电话:(010)64271480
本系统由北京玛格泰克科技发展有限公司设计开发  技术支持:support@magtech.com.cn