Abstract:The spectral responsivity of a high speed, high resolution spectrum measurement system is calibrated via the TG HT9500 high temperature blackbody furnace. The short term stability of the furnace is experimentally studied. A model of the spectral responsivity calibration is established. Calibration is carried out within the temperature range between 1000 to 1800 ℃,the measured spectrum at 1 400 ℃ is compared with the calculation which showed a signal divergence less than 2.7%.
赵云龙,董伟,宦克为,臧延哲. 基于高温黑体的光纤光谱测试系统响应度标定[J]. 计量学报, 2015, 36(6A): 83-85.
ZHAO Yun-long,DONG Wei,HUAN Ke-wei,ZANG Yan-zhe. The Calibration of the Spectral Responsivity of the Fiber Optical Spectrometer Based on High Temperature Blackbody. Acta Metrologica Sinica, 2015, 36(6A): 83-85.