Abstract:An analytical model of the crosstalk for the dual channel of attenuation measurement system is derived through analysis on the principle of a double balanced mixer. An experiment of measuring the crosstalk of a dual channel of attenuation measurement system is carried out. The analytical result is in fare agreement with the measurement data. Based on the analytical result, the isolation of the local oscillator channel should be larger than 120 dB in order to make the uncertainty of the crosstalk less than 0.01 dB at measurement range of 100 dB. The analytical model is of useful value for the establishment of the dual channel of attenuation measurement system and assessment of uncertainty of the crosstalk.
[1]Thompson R, Howes J. The UK national standards of RF and microwave attenuation-a review [C] // British Electromagnetic Measur Conf, London,UK, 2005.
[2]高秋来, 李勇,徐晖. 新建立的衰减计量国家标准装置[J]. 计量学报,2004,25( 1 ): 61-64.
[3]高秋来. 串联替代法衰减测量装置的噪声分析[J]. 计量学报, 2000,21 (4):302-307.
[4]Widarta A, Kawakami T, Suzuki K. Dual channel IF substitution measurement system for microwave attenuation standard [J]. IEICE Transactions on Electronics, 2003,E86-C(8): 1580-1583.
[5]Wu T Y, Shan Y Y, Chua S W. A dual channel AF substitution microwave attenuation measurement system for precision attenuator calibration [C] //Proc of 2007 NCSL International Workshop and Symposium, St Paul, USA,2007.
[6]Wu T Y. Accurate measurement of millimetre-wave attenuation from 75 GHz to 110 GHz using a dual channel heterodyne receiver[J]. Measurement, 2012,45(9) : 1105-1110. [7]Iida H, Widarta A, Kawakami T,et al.Attenuation standard in the frequency range of 50–75 GHz [J]. IEEE Trans Instrum Meas, 2010,59(11): 2921-2929.
[8]Wu T Y. High dynamic range terahertz-wave transmission loss measurement at 330–500 GHz [J].Measurement science and technology, 2012,23(8): 085904.