1. Department of Themal Sicence and Energy Engineering, University of Science & Technology Beijing, Beijing 100083, China
2. National Institute of Metrology, Beijing 100013, China
Abstract:Based on the principle of blackbody radiation,the NIM has been established an irradiation radiation thermometer with center wavelength 900 nm.According to Planck’s law and combining the effective wavelength determination technology in situ, spectral irradiance of a blackbody is measured by this radiation thermometer to calculate the blackbody temperature. The radiation thermometer has the characteristic of simple structure and high stability.The temperature measurement deviation of the thermometry is -0.04 ℃ at the fixed-point of Al with ITS—90 method. And the temperature measurement deviation with RT9032 radiation thermometer is 0.09 ℃ and 0.11 ℃ at the temperature of 800 ℃ and 900 ℃, respectively.
[1]原遵东,赵琪,段宇宁,等.新型精密光电高温计[J].计量学报,2003,24(4):257-261.
[2]Gutschwager B, Schiller S, Hartmann J, et al. A lens-free InGaAs transfer radiation thermometer with an extended calibration range above 960°C[C]//Proceedings TEMPMEKO04.Croatia,2004.
[3]王魁汉.温度测量实用技术[M].北京:机械工业出版社,2006.
[4]原遵东,王铁军,王景辉,等.确定光电高温计的有效波长[J].仪器仪表学报,2007,28(4):265-268.
[5]卢小丰,原遵东,柏成玉,等.基准光电高温计的光谱响应度测量[J]. 计量学报,2007,28(3A):65-68.
[6]Sakuma F, Ma L, Suzuki T, et al. Calibration and characterization of Chino 900 nm silicon narrow-band radiation thermometer[C]//SICE Annual Conference in Sapporo.Sapporo, 2004, 72-76.