肺功能仪用定标筒容量校准方法

任立太,郭立功,包福兵,王金涛,暴雪松

计量学报 ›› 2023, Vol. 44 ›› Issue (1) : 21-25.

PDF(1147 KB)
PDF(1147 KB)
计量学报 ›› 2023, Vol. 44 ›› Issue (1) : 21-25. DOI: 10.3969/j.issn.1000-1158.2023.01.04
力学计量

肺功能仪用定标筒容量校准方法

  • 任立太1,2,郭立功2,包福兵1,王金涛2,暴雪松2
作者信息 +

Method for Volume Calibration of Spirometer Syringes

  • REN Li-tai1,2,GUO Li-gong2,BAO Fu-bing1,WANG Jin-tao2,BAO Xue-song2
Author information +
文章历史 +

摘要

提出一种以静力称重法为基础并结合负压原理的肺功能仪用定标筒(以下称为定标筒)容量校准方法,设计了定标筒容量校准装置。校准时,定标筒活塞杆拉出,定标筒内形成负压,在压力作用下介质(纯水)进入称量筒;采集环境大气压力、介质质量和温度等数据;计算出20℃时定标筒的容量。根据定标筒容量测量原理,建立其容量测量模型。实测某个定标筒,其容量为3.00196L的定标筒,测量结果的不确定度为9.6×10-4 (k=2)。分析表明,该装置具有可靠性,可实现快速校准并满足定标筒容量校准要求

Abstract

A volume calibration method of spirometer syringes (hereinafter referred to as syringes) based on hydro-weighting method and combined with negative pressure principle is presented, and a calibration device for the volume of syringes is designed. During calibration, the medium (pure water) is pumped into the weighing cylinder by the negative pressure formed in syringes when the piston of syringes is pulled out. From the collected data such as ambient atmospheric pressure, medium mass and temperature, the capacity of syringes is calculated at 20°C. The volume measurement model is developed according to the volume measurement principle of syringes. The measured volume of a syringe is 3.00196L, and the uncertainty of measurement result is 9.6×10-4 (k=2). The research shows that the device meets the requirements of syringes capacity calibration and can realize rapid calibration.

关键词

计量学;容量校准方法 / 肺功能仪;静力称重法;负压原理;不确定度

Key words

metrology / volume calibration methods / spirometer syringe / hydro-weighting method / negative pressure principle / uncertainty

引用本文

导出引用
任立太,郭立功,包福兵,王金涛,暴雪松. 肺功能仪用定标筒容量校准方法[J]. 计量学报. 2023, 44(1): 21-25 https://doi.org/10.3969/j.issn.1000-1158.2023.01.04
REN Li-tai,GUO Li-gong,BAO Fu-bing,WANG Jin-tao,BAO Xue-song. Method for Volume Calibration of Spirometer Syringes[J]. Acta Metrologica Sinica. 2023, 44(1): 21-25 https://doi.org/10.3969/j.issn.1000-1158.2023.01.04
中图分类号: TB938.3   

参考文献

[1]Graham B L, Steenbruggen I, Miller M R, et al. Standardization of Spirometry 2019 Update. An Official American Thoracic Society and European Respiratory Society Technical Statement[J]. American Journal of Respiratory and Critical Care Medicine, 2019, 200 (8): e70-e88.
[2]中华医学会, 中华医学会杂志社, 中华医学会全科医学分会, 等. 常规肺功能检查基层指南(2018年)[J]. 中华全科医师杂志, 2019, 18(6):511-518.
[3]中华医学会呼吸病学分会肺功能专业组. 肺功能检查指南(第二部分)—肺量计检查[J]. 中华结核和呼吸杂志, 2014, 37(7):481-486.
[4]郑向真. 肺功能测定中应注意的几个问题[J]. 临床肺科杂志, 2011, 16(10):1619.
Zheng X Z. Several problems that should be noted in the pulmonary function test[J]. Journal of Clinical Pulmonary Medicine, 2011,16(10):1619.
[5]JJF1213-2008 肺功能仪校准规范[S]. 2008.
[6]JJF1586-2016 主动活塞式流量标准装置校准规范[S]. 2016.
[7]JJG18-2008 医用注射器[S]. 2008.
[8]JJG259-2005 标准金属量器[S]. 2005.
[9]Madsen F. Validation of spirometer calibration syringes[J]. Scand J of Clin Lab Invest, 2012, 72(8): 608-613.
[10]吴仲平, 黄锐波, 郑劲平, 等. 肺功能仪容量定标筒的质量检测方法的建立及其应用价值对比研究[J]. 中国全科医学, 2022, 25(2):149-152.
Wu Z P, Huang R B, Zheng J P, et al. Development and application value assessment of a method for measuring the quality of volume calibration syringes[J]. Chinese General Practice, 2022, 25(2):149-152.
[11]Wu Z P, Peng Y Y, Lin K Q, et al. Quality inspection and result analysis of the spirometer calibration cylinder[J]. BMC Pulm Med, 2022, 22: 218.
[12]E542-01 Standard Practice for Calibration of Laboratory Volumetric Apparatus[S].
[13]Picard A, Davis R S, Glaser M, et al. Revised formula for the density of moist air (CIPM-2007) [J]. Metrologia, 2008, 45(2):149-155.
[14]郭立功, 刘子勇, 王金涛, 等. 静力称重法容量测量研究[J]. 计量技术, 2014(12):10-13.
[15]钟家栋, 孙斌, 赵玉晓, 等. 基于静力称重法的微小容量光学检测系统标定[J]. 计量学报, 2020, 41(8): 960-964.
Zhong J D, Sun B, Zhao, Y X, et al. Calibration Method for Optical Measurement of Ultra-Micro Liquid Volume[J]. Acta Metrologica Sinica, 2020, 41(8): 960-964.
[16]Tanaka M, Girard G, Davis R, et al. Recommended table for the density of water between 0℃ and 400℃ based on recent experimental reports [J]. Metrologia, 2001, 38:301-309.
[17]Batista E , Paton R. The Selection of Water Property Formulae for Volume and Flow Calibration [J]. Metrologia, 2007, 44,453-463.
[18]EURAMET cg-19  Guidelines on the Determination of Uncertainty in Gravimetric Volume Calibration[S].
[19]EURAMET cg-18(2015), Guidelines on the Calibration of Non-Automatic Weighing Instruments[S].
[20]郭立功, 刘子勇, 王金涛, 等. 检定介质制备与静力称重法容量测量不确定度研究[J]. 计量学报, 2013, 34(6):534-540.
Guo L G, Liu Z Y, Wang J T, et al. Study on Uncertainty of Volume Measerment by Hydrostatic Weighing Method and Production of Calibration Medium[J]. Acta Metrologica Sinica, 2013,34(6):534-540.
[21]EURAMET cg-21(2013) Guidelines on the Calibration of Standard Capacity Measures Using the Volumetric Method[S].
[22]JJF1059-2012 测量不确定度评定与表示指南[S]. 2012.
[23]陈凌峰. 抽样引起的测量不确定度评定[J]. 计量学报, 2020, 41(7): 891-896.
Chen L F. Evaluating the Sampling Contribution to the Measurement Uncertainty[J]. Acta Metrologica Sinica, 2020, 41(7): 891-896.