高剂量率近距离192Ir治疗源的井型电离室仪器的校准

王继,吴钦宏,赵亮,吕雅竹,王军良,杨小元,王培玮

计量学报 ›› 2021, Vol. 42 ›› Issue (9) : 1232-1236.

PDF(826 KB)
PDF(826 KB)
计量学报 ›› 2021, Vol. 42 ›› Issue (9) : 1232-1236. DOI: 10.3969/j.issn.1000-1158.2021.09.18
电离辐射、标准物质与生物计量

高剂量率近距离192Ir治疗源的井型电离室仪器的校准

  • 王继1,吴钦宏2,赵亮1,吕雅竹1,王军良3,杨小元1,王培玮1
作者信息 +

Calibration of Well Type Ionization Chamber Instruments with High-dose-rate After-loading 192Ir Source

  • WANG Ji1,WU Qin-hong2,ZHAO Liang1,Lü Ya-zhu1,WANG Jun-liang3,YANG Xiao-yuan1,WANG Pei-wei1
Author information +
文章历史 +

摘要

针对后装治疗用192Ir放射源的参考空气比释动能率的量值急需进行溯源。参考国际通例,由指型电离室(PTW-30013)在60Co γ射线和250kV X射线下的空气比释动能校准因子推导得出192Ir γ射线的空气比释动能校准因子,从而测定192Ir放射源参考空气比释动能率标准值,进而完成对井型电离室的参考空气比释动能率校准。通过不确定度评定得出:192Ir放射源的参考空气比释动能率的不确定度为3.6%,井型电离室校准因子的不确定度为3.8%。

Abstract

There is an urgent need for traceability of the reference air kerma rate for the 192Ir source for high-dose-rate after-loading. Referring to the international general practice, the calibration factor of the 192Ir gamma ray is derived from the calibration factor of air kerma of finger ionization chamber under 60Co gamma ray and 250kV X-ray, so as to determine the standard value of the 192Ir gamma ray reference air kerma rate, and then the calibration of reference air kerma rate of well type ionization chamber is completed. The results show that the uncertainty of the 192Ir gamma ray reference air kerma rate is 3.6%, and that of calibration factor of well type ionization chamber is 3.8%.

关键词

计量学;参考空气比释动能率;近距离放射治疗 / 192Ir放射源 / 井型电离室

Key words

metrology / reference air kerma rate / brachytherapy / 192Ir source / well type ionization chamber

引用本文

导出引用
王继,吴钦宏,赵亮,吕雅竹,王军良,杨小元,王培玮. 高剂量率近距离192Ir治疗源的井型电离室仪器的校准[J]. 计量学报. 2021, 42(9): 1232-1236 https://doi.org/10.3969/j.issn.1000-1158.2021.09.18
WANG Ji,WU Qin-hong,ZHAO Liang,Lü Ya-zhu,WANG Jun-liang,YANG Xiao-yuan,WANG Pei-wei. Calibration of Well Type Ionization Chamber Instruments with High-dose-rate After-loading 192Ir Source[J]. Acta Metrologica Sinica. 2021, 42(9): 1232-1236 https://doi.org/10.3969/j.issn.1000-1158.2021.09.18
中图分类号: TB98   

参考文献

[1]Bray F, Ferlay J, Soerjomataram I, et al. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries[J]. CA: A Cancer Journal for Clinicians, 2018, 68(6): 394-424.
[2]刘玮, 罗澜. 引进技术的法制研究——解读《国家中长期科学和技术发展规划纲要(2006—2020年)》[J]. 科技和产业, 2007, 7(8): 42-44.
Liu W, Luo L. Legal System Research About Fetching in Techniques-The explanation of "national medium and long term layout of the development of science and technology program (2006—2020)"[J]. Science Technology and Industry, 2007, 7(8): 42-44.
[3]Kessler C, Kurosawa T, Mikamoto T. Comparison BIPM. RI(I)-K8 of high dose-rate 192Ir brachytherapy standards for reference air kerma rate of the NMIJ and the BIPM[J]. Metrologia, 2014, 52(1A): 06024.
[4]Alvarez J T, De Pooter J A, Andersen C, et al. Comparison BIPM. RI(I)-K8 of high dose-rate 192Ir brachytherapy standards for reference air kerma rate of the VSL and the BIPM[J]. Metrologia, 2014, 51(1A): 06022.
[5]IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy[S].
[6]李兵. 诊断X射线空气比释动能的绝对测量研究[D]. 成都: 成都理工大学, 2013.
[7]吕雅竹,赵瑞,李德红,等.X射线剂量当量仪校准因子不确定度评估[J].  计量学报, 2020, 41(9): 1133-1137.
Lü Y Z, Zhao R, Li D H, et al. Uncertainty Evaluation of Calibration Factor for X-ray Dose Equivalent Instrument[J]. Acta Metrologica Sinica,2020, 41(9): 1133-1137.
[8]王培玮,郑伟宁. 用于同步辐射X射线注量绝对测量的自由空气电离室[J].  计量学报, 2020, 41(10): 1303-1307.
Wang P W, Zheng W N. Free Air Ionization Chamber for the Absolute Measurement of Synchrotron Radiation X-ray Fluence[J]. Acta Metrologica Sinica,2020, 41(10): 1303-1307.
[9]罗素明, 何志坚, 高磊. 测量后装近距离放射治疗192Ir源参考空气比释动能方法学研究[J]. 中华放射医学与防护杂志, 2008, 28(4): 398-400.
Luo S M, He Z J, Gao L. Development of measurement methodology with reference air kerma of 192Ir source afterloding in bracytherapy[J]. Chin J Radiol Med Prot, 2008, 28(4): 398-400.
[10]Goetsch, Steven J. Calibration of 192Ir high-dose-rate afterloading systems[J]. Medical Physics, 1991, 18(3): 462.
[11]Dijk E V, Kolkman-Deurloo I K K, Damen P M G. Determination of the reference air kerma rate for 192Ir brachytherapy sources and the related uncertainty[J]. Medical Physics, 2004, 31(10): 2826-2833.
[12]陈凌峰. 抽样引起的测量不确定度评定[J].  计量学报, 2020, 41(7): 891-896.
Chen L F. Evaluating the Sampling Contribution to the Measurement Uncertainty[J]. Acta Metrologica Sinica,2020, 41(7): 891-896.
[13]陈怀艳, 曹芸, 韩洁, 等. 测量不确定度的发展和应用研究[J]. 宇航计测技术, 2014,(5): 65-70.
Chen H Y, Cao Y, Han J, et al. Research on Development and Application of Measurement Uncertainty[J]. Journal of Astronautic Metrology and Measurement, 2014,(5): 65-70.

PDF(826 KB)

Accesses

Citation

Detail

段落导航
相关文章

/

〈 〉