反宇宙射线低本底γ谱仪系统的设计和建立

张朕朕,范梓浩,周倩倩,梁珺成,张明

计量学报 ›› 2020, Vol. 41 ›› Issue (12A) : 167-171.

PDF(1225 KB)
PDF(1225 KB)
计量学报 ›› 2020, Vol. 41 ›› Issue (12A) : 167-171. DOI: 10.3969/j.issn.1000-1158.2020.Z1.30
电离辐射、标准物质与生物计量

反宇宙射线低本底γ谱仪系统的设计和建立

  • 张朕朕1,范梓浩1,2,周倩倩1,梁珺成1,张明1
作者信息 +

Design and Establishment of Anti-cosmic Ray Low Background Gamma Spectrometer System

  • ZHANG Zhen-zhen1,FAN Zi-hao1,2,ZHOU Qian-qian1,LIANG Jun-cheng1,ZHANG Ming1
Author information +
文章历史 +

摘要

通过对γ谱仪中的本底来源和屏蔽方法的研究,建立了一套反宇宙射线低本底γ谱仪测量系统,100~2500keV能区积分本底为1.1×10-1s-1/100cm3Ge。应用该系统测量几何规格为75mm×70mm土壤基材和75mm×35mm生物灰基材的标准样品,确定了土壤样品中137Cs的最小可探测活度为0.089Bq;生物灰样品中137Cs 和134Cs的最小可探测活度分别为0.068Bq和0.098Bq。

Abstract

Based on the study of background and shielding method in gamma spectrometer, an anti-cosmic low-background γ-ray spectrometer system which the integrated background is up to 1.1×10-1s-1/100cm3Ge in the energy region between 100keV and 2500keV was built. Apply this device to measure the standard samples of 75mm×70mm soil substrate and 75mm×35mm biological ash substrate. The minimum detectable activity of137 cs in soil samples was 0.089Bq; the minimum detectable activity of137 Cs and134 Cs in biological gray samples were 0.068Bq and 0.098Bq.

关键词

计量学 / 低本底&gamma / 谱仪 / 放射性核素 / 最小可探测活度

Key words

metrology / low background Gamma spectrometer / radionuclide / minimum detectable activity

引用本文

导出引用
张朕朕,范梓浩,周倩倩,梁珺成,张明. 反宇宙射线低本底γ谱仪系统的设计和建立[J]. 计量学报. 2020, 41(12A): 167-171 https://doi.org/10.3969/j.issn.1000-1158.2020.Z1.30
ZHANG Zhen-zhen,FAN Zi-hao,ZHOU Qian-qian,LIANG Jun-cheng,ZHANG Ming. Design and Establishment of Anti-cosmic Ray Low Background Gamma Spectrometer System[J]. Acta Metrologica Sinica. 2020, 41(12A): 167-171 https://doi.org/10.3969/j.issn.1000-1158.2020.Z1.30
中图分类号: TB98   

参考文献

[1]范梓浩. 反宇宙射线超低本底伽马谱仪的研究设计[D]. 南昌: 东华理工大学, 2018.
[2]Arnold J R, Libby W F. Age Determinations by Radiocarbon Content: Checks with Samples of Known Age[J]. Science, 1949, 110(2869): 678-680.
[3]Hutchinson J M R, Mann W B, Perkins R W. Low-level radioactivity measurements[J]. Nuclear Instruments & Methods, 1973, 112(1-2): 305-318.
[4]Heusser G, Weber M, Hakenmüller J, et al. GIOVE: a new detector setup for high sensitivity germanium spectroscopy at shallow depth[J]. The European Physical Journal C, 2015, 75(11): 1-16.
[5]Heusser G, Weber M, Denz T, et al. GIOVE, a shallow laboratory Ge-spectrometer with 100 μBq/kg sensitivity[C]// AIP Conference Proceedings. 2013: 12-15.
[6]刁立军, 侯铁栋, 姚顺和, 等. 新建反宇宙射线低本底γ谱仪及其应用[C]//全国核电子学与核探测技术学术年会. 2010: 526-530.
[7]Laurec J, Blanchard X, Pointurier F, et al. A new low background gamma spectrometer equipped with an anti-cosmic device[J]. Nuclear Instruments and Methods in Physics Research, 1996, 369(2-3): 566-571.
[8]Berger M J, Hubbell J H, et al. XCOM: Photon Cross Section Database (version 1.5)[EB/OL]. http://physics.nist.gov/xcom
[9]Benettoni M, Bonomi G, Calvini P, et al. Muon radiography with the CMS Muon Barrel chambers[C]// IEEE. Nuclear Science Symposium Conference Record. 2008: 1021-1025.
[10]侯铁栋. 反宇宙射线低本底γ谱仪本底来源及降低方法[D]. 南宁: 广西大学, 2010.
[11]Annunziata M. Handbook of Radioactivity Anlysis Second Edition[M]. Elsevier/Academic Press, 2003.

基金

质量技术监督能力提升写项目(25-ANL1819)