环道式气体流量标准装置的不确定度评估及能力验证

李春辉,崔骊水,王池,曹鹏,王粟

计量学报 ›› 2020, Vol. 41 ›› Issue (12A) : 7-10.

PDF(1185 KB)
PDF(1185 KB)
计量学报 ›› 2020, Vol. 41 ›› Issue (12A) : 7-10. DOI: 10.3969/j.issn.1000-1158.2020.Z1.02
流量计量

环道式气体流量标准装置的不确定度评估及能力验证

  • 李春辉1,崔骊水1,王池1,曹鹏1,王粟2
作者信息 +

Uncertainty Evaluation and Capability Verification of Gas Flow Close Loop Standard Facility

  • LI Chun-hui1,CI Li-shui1,WANG Chi1,CAO Peng1,WANG Su2
Author information +
文章历史 +

摘要

2014年底,中国计量科学研究院(NIM)建立了一套环道式气体流量标准装置,该标准装置采用4台DN100的涡轮流量计作为标准表,该装置在压力190~2500kPa内,可检测的流量范围为40~1400m3/h。2020年1月, 环道式气体流量标准装置通过建标考核,2020年3月对外正式开展检测服务。基于标准表的原位校准技术,及绝压+差压的压力测量方法,标准装置可校准流量计的最佳不确定度为0.18%(k=2)。

Abstract

At the end of 2014, the first gas flow close loop standard facility in China was built in National Institute of Metrology (NIM). The standard facility uses four DN100 turbine flowmeters as master meters, and the achievable flowrate range is 40~1400m3/h as well as the pressure range is 190~2500kPa. Based on the in-situ calibration technology of the master meter and the pressure measurement method of absolute pressure+differential pressure, the best uncertainty of the standard facility to calibrate the flowmeter is 0.18%(k=2). In January 2020, the gas flow close loop standard facility passed the construction standard assessment, and the testing service was officially launched in March 2020.

关键词

计量学 / 气体流量 / 环道式气体流量标准装置 / 不确定度评估

Key words

metrology / gas flow / gas flow close loop standard facility / uncertainty evaluation

引用本文

导出引用
李春辉,崔骊水,王池,曹鹏,王粟. 环道式气体流量标准装置的不确定度评估及能力验证[J]. 计量学报. 2020, 41(12A): 7-10 https://doi.org/10.3969/j.issn.1000-1158.2020.Z1.02
LI Chun-hui,CI Li-shui,WANG Chi,CAO Peng,WANG Su. Uncertainty Evaluation and Capability Verification of Gas Flow Close Loop Standard Facility[J]. Acta Metrologica Sinica. 2020, 41(12A): 7-10 https://doi.org/10.3969/j.issn.1000-1158.2020.Z1.02
中图分类号: TB937   

参考文献

[1]BP Statistical Review of Word Energy[EB]. https://www.bp.com/en/global/corporate/energy-economics/statistical-review-of-world-energy.html.[2020-08-19].
[2]Wright J D, Mickan B, Vallet J P, et al. Extended data analysis of bilateral comparisons with air and natural gas up to 5 MPa[C]//Flomeko, Sydney Australia, 2016.
[3]Li C H, Mickan B, Cui L S, et al. The high pressure sonic nozzle gas flow standard facility in NIM[C]//Flomeko, Lisbon, Portuguesa, 2019.
[4]Bremser W, Hasselbarth W, Hirlehei U, et al. Uncertainty analysis and long-term stability investigation of the German primary high-pressure natural gas test facility Pigsar[C]//Flomeko, Groningen, 2003.
[5]苏一林. 基于高压音速喷嘴法气体流量标准装置的涡轮流量计计量性能研究[D]. 保定: 河北大学, 2017.
[6]Cox M G. The Evaluation of key comparison data[J]. Metrologia, 2002, 39(6): 589-595.
[7]Cox M G. The evaluation of key comparison data: determining the largest consistent subset[J]. Metrologia, 44(3): 187-200, 2007.

基金

国家重点研发计划(2017YFF0205305)