1.College of Engineering Measurement, China Jiliang University, Hangzhou, Zhejiang 310018, China
2.National Institute of Metrology, Beijing 100029, China
3.Hebei University, Baoding, Hebei 100075, China
Abstract:The high pressure close loop gas flow standard facility in National Insititute of Metrology is discribed, which is constructed in 2014. The introduction and uncertainty analysis of the facility was investigated. With 4 sets of turbine meters as the master meters, the maximum flowrate could be 1600m3/h at 2.5MPa for the high pressure close loop working gas flow standard. The expanded uncertainty is 0.21% (k=2). The uncertainty of the facility realization was guaranteed by the calibration of master flowmeters in original position and the pressure measurement method with absolute pressure transmitters plus differential pressure transmitters.
张翰,李春辉,崔骊水,曹培娟. 高压环道气体流量标准装置[J]. 计量学报, 2017, 38(3): 324-327.
ZHANG Han,LI Chun-hui,CUI Li-shui,CAO Pei-juan. The Close Loop Standard Facility of High Pressure Gas Flow. Acta Metrologica Sinica, 2017, 38(3): 324-327.
[1]Mickan B, Kramer R, Dopheide D. The linking of the CIPM key comparisons CCM.FF-KC5a (for natural gas) and CCM.FF-KC5b (compressed air/nitrogen) using model based analysis of the data[C]//14th International Flow Measurement Conference on Flow Measurement, Sandton, Gauteng, South Afica, 2007: 281-302.
[2]曹培娟, 李春辉, 崔骊水, 等. 正压法pVTt气体流量标准装置的建立[J]. 计量技术, 2015, (9): 42-44.
[3]李春辉,李鹏. 基于雷诺数对涡轮流量计仪表特性评估方法探析[J]. 计量学报, 2014, 35(6A): 150-153.