Abstract:Based on laser interference technology and free-fall theory, the National Institute of Metrology (China) has developed the latest NIM-3A type absolute gravimeter.This instrument has participated in the 9th international comparison of absolute gravimetries CCM.G-K2,according to the published result of CCM.G-K2,the NIM-3A type absolute gravimeter with the combined standard uncertainty of 5.3 μGal is agreed with the key comparison reference value and the degree of equivalence is gained.
[1]Jiang Z,Pálinká? V,Arias F E , et al. The 8th International Comparison of Absolute Gravimeters 2009: the first Key Comparison (CCM.G-K1) in the field of absolute gravimetry[J]. Metrologia,2012,49(6): 666.
[2]吴书清,葛春雨,刘达伦,等.基于FPGA的重力仪干涉条纹和时间间隔计数[J].计量学报,2010,31(6):530-532.
[3]吴书清,李春剑,粟多武,等.绝对重力仪国际比对新动态[J].计量学报,2013,34(6):545-547.
[4]Francis O,Baumann H,Ullrich C,et al.CCM.G-K2 key comparison[J].Metrologia,2015,52:07009.