Abstract:The design and development technique of a WR-15 microcalorimeter based on a thermistor mount is presented. The microcalorimeter is designed as twn-line structure which can lower the influnece from the laboratory temperature floating. The basic principle of the measurement is presented too. The microcalorimeter will be national primary standard of WR-15 (50~75 GHz) millimeter-wave power. It enhanced China RF power measurement capability up to 75 GHz.
崔孝海, 李勇, 张淑溢, 孙伟杰. 毫米波WR-15(50~75 GHz)国家功率基准的研制[J]. 计量学报, 2014, 35(4): 378-381.
CUI Xiao-hai, LI Yong, ZHANG Shu-yi, SUN Wei-jie. Design and Development of the WR-15 (50~75) GHz Millimeter-wave Power National Primary Standard. Acta Metrologica Sinica, 2014, 35(4): 378-381.
[1] Crowley T P, Cui X H.Design and evaluation of a WR-15(50 to 75 GHz) microcalorimeter[C]// CPEM 2008 Conference,Broomfield,Colorado,USA, 2008.
[2] Allen J W, Fred R C, Larsen N T, et al. The NIST Microwave Power Standards in Waveguide[R]. NIST Technical Note 1511, 1999.
[3] Cui X H, Crowley T P.Comparison of Experimental Techniques for Evaluating the Correction Factor of a Rectangular Waveguide Microcalorimeter[C]//IEEE Transactions on Instrumentation and Measurement,Daejeon, Korea,2010.