Abstract:To investigate the static pressure coefficient and temperature coefficient of the laboratory standard microphones, a static pressure and temperature control system was developed based on the coupler reciprocity system. The sensitivities under different pressure and temperature conditions were dealt with application of Straight Line Fitting Method and the pressure and temperature coefficient at different frequency was obtained. The results show that in low frequency the coefficients hold the line, and at half of the resonant frequency of diaphragm, the pressure coefficient gets its max., while the temperature coefficient gets its min.
何龙标,白滢,裘剑敏,钟波,牛锋. 实验室标准传声器静压和温度修正系数的测量[J]. 计量学报, 2012, 33(4): 336-339.
HE LONG-biao,BAI Ying,QIU Jian-min,ZHONG Bo,NIU Feng. Measurement of the Pressure and Temperature Coefficient of Laboratory Standard Microphone. Acta Metrologica Sinica, 2012, 33(4): 336-339.
[1] IEC 61094-1,Measurement microphones Part 1:Specifications for laboratory standard microphones[S].1995.
[2]IEC 61094-2,Measurement microphones Part 2: Primary method for pressure calibration of laboratory standard microphones by the reciprocity technique [S]. 2009.
[3]Rasmussen K.Final report on the key comparison CCAUV.A-K3[R].Vicente Cntanda Henriquez(Prosent address:University of Vigo,Spain),CENAM/DPLA, 2005.
[4]何龙标,许欢,牛锋,等. 实验室标准传声器前腔体积的测量方法研究[J],计量学报,2010,31(2):97-100.
[5]Ryuzo Horiuchi, Takeshi Fujimori, Sojun Sato.Uncertainty analysis for pressure sensitivities of laboratory standard microphones[J].Acoustical Science and Technology-- Acoustical Science of Japan,2004,25(5):354-363.